CVE-2017-12712
highpriority · track · 35EPSS 0.010 · 64 %ileEUUSEPSS · scored 2026-10-27
The authentication algorithm in Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017, which involves an authentication key and time stamp, can be compromised or bypassed, which may allow a nearby attacker to issue unauthorized commands to the pacemaker via RF communications. CVSS v3 base score: 7.5, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
remediations
Vendor fix, mitigation, or workaround guidance, as published by each source.
A vendor fix is available.
- FixedCurrentcisa_csaf_ot
Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.
- FixedEllipsecisa_csaf_ot
Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.
- FixedFortifycisa_csaf_ot
Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.
- FixedFortify Assuracisa_csaf_ot
Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.
- FixedPromotecisa_csaf_ot
Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.
- FixedPromote Quadracisa_csaf_ot
Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.
- FixedQuadra Assuracisa_csaf_ot
Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.
- FixedQuadra Assura MPcisa_csaf_ot
Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.
- FixedUnifycisa_csaf_ot
Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.
- FixedUnify Assuracisa_csaf_ot
Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.
- FixedUnify Quadracisa_csaf_ot
Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.
- MitigationAccent MRIcisa_csaf_ot
The pacemaker firmware update will implement RF wake-up protections and limit the commands that can be issued to pacemakers via RF communications. Additionally the updated pacemaker firmware will prevent unencrypted transmission of patient information (Accent and Anthem only). The firmware update can be applied to an implanted pacemaker via the Merlin PCS Programmer by a healthcare provider. It is recommended that healthcare providers discuss this update with their patients and carefully consider the potential risk of a cybersecurity attack along with the risk of performing a firmware update. Implementation of the firmware update is to be determined based on the physician's professional judgment and patient management considerations. Pacemakers manufactured beginning August 28, 2017, will have this update preloaded on devices.
- MitigationAccent/Anthemcisa_csaf_ot
The pacemaker firmware update will implement RF wake-up protections and limit the commands that can be issued to pacemakers via RF communications. Additionally the updated pacemaker firmware will prevent unencrypted transmission of patient information (Accent and Anthem only). The firmware update can be applied to an implanted pacemaker via the Merlin PCS Programmer by a healthcare provider. It is recommended that healthcare providers discuss this update with their patients and carefully consider the potential risk of a cybersecurity attack along with the risk of performing a firmware update. Implementation of the firmware update is to be determined based on the physician's professional judgment and patient management considerations. Pacemakers manufactured beginning August 28, 2017, will have this update preloaded on devices.
- MitigationAssurity MRIcisa_csaf_ot
The pacemaker firmware update will implement RF wake-up protections and limit the commands that can be issued to pacemakers via RF communications. Additionally the updated pacemaker firmware will prevent unencrypted transmission of patient information (Accent and Anthem only). The firmware update can be applied to an implanted pacemaker via the Merlin PCS Programmer by a healthcare provider. It is recommended that healthcare providers discuss this update with their patients and carefully consider the potential risk of a cybersecurity attack along with the risk of performing a firmware update. Implementation of the firmware update is to be determined based on the physician's professional judgment and patient management considerations. Pacemakers manufactured beginning August 28, 2017, will have this update preloaded on devices.
- MitigationAssurity/Allurecisa_csaf_ot
The pacemaker firmware update will implement RF wake-up protections and limit the commands that can be issued to pacemakers via RF communications. Additionally the updated pacemaker firmware will prevent unencrypted transmission of patient information (Accent and Anthem only). The firmware update can be applied to an implanted pacemaker via the Merlin PCS Programmer by a healthcare provider. It is recommended that healthcare providers discuss this update with their patients and carefully consider the potential risk of a cybersecurity attack along with the risk of performing a firmware update. Implementation of the firmware update is to be determined based on the physician's professional judgment and patient management considerations. Pacemakers manufactured beginning August 28, 2017, will have this update preloaded on devices.
cvss
Technical severity, 0–10. One row per scoring source.
base score 0–10 · higher = more severe · hover a vector to decode
weaknesses
The CWE weakness classes this flaw belongs to.
affected products (34)
Products and version ranges known to be affected.
advisories (2)
Vendor and ecosystem advisories describing this.
- ICSMA-17-241-01ICSMA-17-241-01_Abbott Laboratories ' Accent/Anthem, Accent MRI, Assurity/Allure, and Assurity MRI Pacemaker Vulnerabilities2017-08-29icsa · cisa_csaf_otUS
This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov
- ICSMA-18-107-01Abbott Laboratories Defibrillator2018-04-17icsa · cisa_csaf_otUS
Successful exploitation of these vulnerabilities may allow a nearby attacker to gain unauthorized access to an ICD to issue commands, change settings, or otherwise interfere with the intended function of the ICD.
also known as (3)
Other names and identifiers for this entity.
sources
Which sources assert this record, and when.
references (7)
External citations — patches, advisories, analysis.
- http://abbott.mediaroom.com/2017-08-29-Abbott-issues-new-updates-for-implanted-cardiac-devicesPatchcnnvd
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01cnnvd
- https://www.securityfocus.com/bid/100523cnnvd
- http://www.securityfocus.com/bid/100523enisa_euvd
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01enisa_euvd
- http://www.securityfocus.com/bid/100523Third Party AdvisoryVDB Entrynvd
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01Third Party AdvisoryUS Government Resourcenvd