ICSMA-17-241-01

icsaUScisa_csaf_ot

ICSMA-17-241-01_Abbott Laboratories ' Accent/Anthem, Accent MRI, Assurity/Allure, and Assurity MRI Pacemaker Vulnerabilities

view advisory published 2017-08-29

This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov

01

describes (3)

Related CVEs.

IDseverity exploitable
CVE-2017-12716USmedium6.5OT elevated—
Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitorin…
CVE-2017-12714USmedium6.5—
Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted "RF wake-up" commands that can be received, which may a…
CVE-2017-12712UShigh8.8—
The authentication algorithm in Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017, which involves an authentication key and time stamp, can be compromised or bypass…
02

references

External citations — patches, advisories, analysis.

03

remediations (12)

  • MitigationAccent MRI

    The pacemaker firmware update will implement RF wake-up protections and limit the commands that can be issued to pacemakers via RF communications. Additionally the updated pacemaker firmware will prevent unencrypted transmission of patient information (Accent and Anthem only). The firmware update can be applied to an implanted pacemaker via the Merlin PCS Programmer by a healthcare provider. It is recommended that healthcare providers discuss this update with their patients and carefully consider the potential risk of a cybersecurity attack along with the risk of performing a firmware update. Implementation of the firmware update is to be determined based on the physician's professional judgment and patient management considerations. Pacemakers manufactured beginning August 28, 2017, will have this update preloaded on devices.

  • MitigationAccent MRI

    The pacemaker firmware update will implement RF wake-up protections and limit the commands that can be issued to pacemakers via RF communications. Additionally the updated pacemaker firmware will prevent unencrypted transmission of patient information (Accent and Anthem only). The firmware update can be applied to an implanted pacemaker via the Merlin PCS Programmer by a healthcare provider. It is recommended that healthcare providers discuss this update with their patients and carefully consider the potential risk of a cybersecurity attack along with the risk of performing a firmware update. Implementation of the firmware update is to be determined based on the physician's professional judgment and patient management considerations. Pacemakers manufactured beginning August 28, 2017, will have this update preloaded on devices.

  • MitigationAccent MRI

    The pacemaker firmware update will implement RF wake-up protections and limit the commands that can be issued to pacemakers via RF communications. Additionally the updated pacemaker firmware will prevent unencrypted transmission of patient information (Accent and Anthem only). The firmware update can be applied to an implanted pacemaker via the Merlin PCS Programmer by a healthcare provider. It is recommended that healthcare providers discuss this update with their patients and carefully consider the potential risk of a cybersecurity attack along with the risk of performing a firmware update. Implementation of the firmware update is to be determined based on the physician's professional judgment and patient management considerations. Pacemakers manufactured beginning August 28, 2017, will have this update preloaded on devices.

  • MitigationAccent/Anthem

    The pacemaker firmware update will implement RF wake-up protections and limit the commands that can be issued to pacemakers via RF communications. Additionally the updated pacemaker firmware will prevent unencrypted transmission of patient information (Accent and Anthem only). The firmware update can be applied to an implanted pacemaker via the Merlin PCS Programmer by a healthcare provider. It is recommended that healthcare providers discuss this update with their patients and carefully consider the potential risk of a cybersecurity attack along with the risk of performing a firmware update. Implementation of the firmware update is to be determined based on the physician's professional judgment and patient management considerations. Pacemakers manufactured beginning August 28, 2017, will have this update preloaded on devices.

  • MitigationAccent/Anthem

    The pacemaker firmware update will implement RF wake-up protections and limit the commands that can be issued to pacemakers via RF communications. Additionally the updated pacemaker firmware will prevent unencrypted transmission of patient information (Accent and Anthem only). The firmware update can be applied to an implanted pacemaker via the Merlin PCS Programmer by a healthcare provider. It is recommended that healthcare providers discuss this update with their patients and carefully consider the potential risk of a cybersecurity attack along with the risk of performing a firmware update. Implementation of the firmware update is to be determined based on the physician's professional judgment and patient management considerations. Pacemakers manufactured beginning August 28, 2017, will have this update preloaded on devices.

  • MitigationAccent/Anthem

    The pacemaker firmware update will implement RF wake-up protections and limit the commands that can be issued to pacemakers via RF communications. Additionally the updated pacemaker firmware will prevent unencrypted transmission of patient information (Accent and Anthem only). The firmware update can be applied to an implanted pacemaker via the Merlin PCS Programmer by a healthcare provider. It is recommended that healthcare providers discuss this update with their patients and carefully consider the potential risk of a cybersecurity attack along with the risk of performing a firmware update. Implementation of the firmware update is to be determined based on the physician's professional judgment and patient management considerations. Pacemakers manufactured beginning August 28, 2017, will have this update preloaded on devices.

  • MitigationAssurity MRI

    The pacemaker firmware update will implement RF wake-up protections and limit the commands that can be issued to pacemakers via RF communications. Additionally the updated pacemaker firmware will prevent unencrypted transmission of patient information (Accent and Anthem only). The firmware update can be applied to an implanted pacemaker via the Merlin PCS Programmer by a healthcare provider. It is recommended that healthcare providers discuss this update with their patients and carefully consider the potential risk of a cybersecurity attack along with the risk of performing a firmware update. Implementation of the firmware update is to be determined based on the physician's professional judgment and patient management considerations. Pacemakers manufactured beginning August 28, 2017, will have this update preloaded on devices.

  • MitigationAssurity MRI

    The pacemaker firmware update will implement RF wake-up protections and limit the commands that can be issued to pacemakers via RF communications. Additionally the updated pacemaker firmware will prevent unencrypted transmission of patient information (Accent and Anthem only). The firmware update can be applied to an implanted pacemaker via the Merlin PCS Programmer by a healthcare provider. It is recommended that healthcare providers discuss this update with their patients and carefully consider the potential risk of a cybersecurity attack along with the risk of performing a firmware update. Implementation of the firmware update is to be determined based on the physician's professional judgment and patient management considerations. Pacemakers manufactured beginning August 28, 2017, will have this update preloaded on devices.

  • MitigationAssurity MRI

    The pacemaker firmware update will implement RF wake-up protections and limit the commands that can be issued to pacemakers via RF communications. Additionally the updated pacemaker firmware will prevent unencrypted transmission of patient information (Accent and Anthem only). The firmware update can be applied to an implanted pacemaker via the Merlin PCS Programmer by a healthcare provider. It is recommended that healthcare providers discuss this update with their patients and carefully consider the potential risk of a cybersecurity attack along with the risk of performing a firmware update. Implementation of the firmware update is to be determined based on the physician's professional judgment and patient management considerations. Pacemakers manufactured beginning August 28, 2017, will have this update preloaded on devices.

  • MitigationAssurity/Allure

    The pacemaker firmware update will implement RF wake-up protections and limit the commands that can be issued to pacemakers via RF communications. Additionally the updated pacemaker firmware will prevent unencrypted transmission of patient information (Accent and Anthem only). The firmware update can be applied to an implanted pacemaker via the Merlin PCS Programmer by a healthcare provider. It is recommended that healthcare providers discuss this update with their patients and carefully consider the potential risk of a cybersecurity attack along with the risk of performing a firmware update. Implementation of the firmware update is to be determined based on the physician's professional judgment and patient management considerations. Pacemakers manufactured beginning August 28, 2017, will have this update preloaded on devices.

  • MitigationAssurity/Allure

    The pacemaker firmware update will implement RF wake-up protections and limit the commands that can be issued to pacemakers via RF communications. Additionally the updated pacemaker firmware will prevent unencrypted transmission of patient information (Accent and Anthem only). The firmware update can be applied to an implanted pacemaker via the Merlin PCS Programmer by a healthcare provider. It is recommended that healthcare providers discuss this update with their patients and carefully consider the potential risk of a cybersecurity attack along with the risk of performing a firmware update. Implementation of the firmware update is to be determined based on the physician's professional judgment and patient management considerations. Pacemakers manufactured beginning August 28, 2017, will have this update preloaded on devices.

  • MitigationAssurity/Allure

    The pacemaker firmware update will implement RF wake-up protections and limit the commands that can be issued to pacemakers via RF communications. Additionally the updated pacemaker firmware will prevent unencrypted transmission of patient information (Accent and Anthem only). The firmware update can be applied to an implanted pacemaker via the Merlin PCS Programmer by a healthcare provider. It is recommended that healthcare providers discuss this update with their patients and carefully consider the potential risk of a cybersecurity attack along with the risk of performing a firmware update. Implementation of the firmware update is to be determined based on the physician's professional judgment and patient management considerations. Pacemakers manufactured beginning August 28, 2017, will have this update preloaded on devices.