ICSMA-18-107-01

icsaUScisa_csaf_ot

Abbott Laboratories Defibrillator

view advisory published 2018-04-17

Successful exploitation of these vulnerabilities may allow a nearby attacker to gain unauthorized access to an ICD to issue commands, change settings, or otherwise interfere with the intended function of the ICD.

01

describes (2)

Related CVEs.

IDseverity exploitable
CVE-2017-12714USmedium6.5—
Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted "RF wake-up" commands that can be received, which may a…
CVE-2017-12712UShigh8.8—
The authentication algorithm in Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017, which involves an authentication key and time stamp, can be compromised or bypass…
02

references

External citations — patches, advisories, analysis.

03

remediations (22)

  • FixedCurrent

    Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.

  • FixedCurrent

    Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.

  • FixedEllipse

    Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.

  • FixedEllipse

    Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.

  • FixedFortify

    Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.

  • FixedFortify

    Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.

  • FixedFortify Assura

    Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.

  • FixedFortify Assura

    Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.

  • FixedPromote

    Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.

  • FixedPromote

    Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.

  • FixedPromote Quadra

    Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.

  • FixedPromote Quadra

    Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.

  • FixedQuadra Assura

    Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.

  • FixedQuadra Assura

    Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.

  • FixedQuadra Assura MP

    Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.

  • FixedQuadra Assura MP

    Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.

  • FixedUnify

    Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.

  • FixedUnify

    Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.

  • FixedUnify Assura

    Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.

  • FixedUnify Assura

    Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.

  • FixedUnify Quadra

    Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.

  • FixedUnify Quadra

    Abbott 's older generation devices (i.e., Current and Promote) are not capable of accepting the firmware update due to technology limitations. If healthcare providers and patients have any concerns relating to device cybersecurity for those patients implanted with Current/Promote devices, providers have the option to permanently disable the RF communication capability in the device. However, if this option is selected, the patient can no longer be monitored remotely using an RF Merlin@home transmitter. For most patients, permanently disabling RF is not advisable given the proven benefits and improved survival associated with home monitoring.