CVE-2020-25175
mediumpriority · track · 39EPSS 0.011 · 67 %ileEUUSEPSS · scored 2026-10-27
GE Healthcare Imaging and Ultrasound Products may allow specific credentials to be exposed during transport over the network.
remediations
Vendor fix, mitigation, or workaround guidance, as published by each source.
A vendor fix is available.
- Fixed1.5T Brivo MR355cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- Fixed1.5T Signa Hdxcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- Fixed3.0T Signa HDXxcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- Fixed3.0T Signa Hdxtcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedAMX 700cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedAWcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedAWScisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedAlaris PC Unitcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedBrightSpeed Edgecisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedBrightSpeed Edge Selectcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedBrightSpeed Elitecisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedBrightSpeed Elite Selectcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedBrivio NM 615cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedBrivo CT385cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedBrivo XR118cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedBrivo XR383cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedBrivo XR515cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedBrivo XR575cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedDefinium 5000cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedDefinium 6000cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedDefinium 8000cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedDiscovery CT590RTcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedDiscovery CT750HDcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedDiscovery IQcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedDiscovery MI MI DRcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedDiscovery NM 630cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedDiscovery NM 750bcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedDiscovery NM D530ccisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedDiscovery NM/CT 670cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedDiscovery NM/CT 850cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedDiscovery NM/CT 860cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedDiscovery NM/CT 870cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedDiscovery NM/CT D570ccisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedDiscovery NM830cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedDiscovery XR650cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedDiscovery XR656cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedDiscovery XR656+cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedEchoPAC (Turnkey)cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedEssentialcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedGE Imaging and Ultrasound Productscisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedImage Vault (Turnkey)cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedInfiniacisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedInnova 2000cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedInnova 2100-IQcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedInnova 212-IQcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedInnova 3100cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedInnova 3100-IQcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedInnova 313-IQcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedInnova 4100cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedInnova 4100-IQcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedInnova IGS 5x0cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedInnova IGS 6x0cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedInnova IGS 7x0cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedLOGIQ 5cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedLOGIQ 7cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedLOGIQ 9cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedLightSpeed Pro16cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedLightSpeed RT16cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedLightSpeed VCTcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedOptima 3100cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedOptima 320cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedOptima Advancecisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedOptima CL320cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedOptima CL320icisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedOptima CL323icisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedOptima CT520cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedOptima CT540cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedOptima CT580cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedOptima CT580RTcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedOptima CT580Wcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedOptima CT660cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedOptima CT670cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedOptima CT680 Expertcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedOptima CT680 Professionalcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedOptima CT680 Quantumcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedOptima IGS 320cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedOptima IGS 330cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedOptima MR360cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedOptima NM/CT640cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedOptima XR200amxcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedOptima XR220amxcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedOptima XR640cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedOptima XR646cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedPET Discovery IQcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedPET Discovery IQ upgradecisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedPETrace 800cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedPrecision 500Dcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedPrecision WDR1cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedRevolution ACTscisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedRevolution CTcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedRevolution Discover CTcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedRevolution EVOcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedRevolution Frontiercisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedRevolution Frontier EScisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedRevolution HACTcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedRevolution HDcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedSeno 200Dcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedSeno DScisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedSenographe Pristinacisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedSigna HDicisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedSigna VIBRANTcisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedVentricisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedVivid 7cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedVivid Icisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedVoluson 730cisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
- FixedXeleriscisa_csaf_ot
In addition to the product specific recommendations, GE recommends utilizing clinical network security best practices. Suggested clinical best practices include:
cvss
Technical severity, 0–10. One row per scoring source.
base score 0–10 · higher = more severe · hover a vector to decode⚠ sources disagree · medium–critical · 5–9.8
weaknesses
The CWE weakness classes this flaw belongs to.
affected products (331)
Products and version ranges known to be affected.
advisories (1)
Vendor and ecosystem advisories describing this.
- ICSMA-20-343-01GE Healthcare Imaging and Ultrasound Products2020-12-08icsa · cisa_csaf_otUS
Successful exploitation of these vulnerabilities could occur if an attacker gains access to the healthcare delivery organization 's (HDO) network. If exploited, these vulnerabilities could allow an attacker to gain access to affected devices in a way that is comparable with GE (remote) service user privileges. A successful exploitation could expose sensitive data such as a limited set of patient health information (PHI) or could allow the attacker to run arbitrary code, which might impact the availability of the system and allow manipulation of PHI.
also known as (3)
Other names and identifiers for this entity.
sources
Which sources assert this record, and when.
references (6)
External citations — patches, advisories, analysis.
- https://nvd.nist.gov/vuln/detail/CVE-2020-25175cnnvd
- https://us-cert.cisa.gov/ics/advisories/icsma-20-343-01cnnvd
- https://www.auscert.org.au/bulletins/ESB-2020.4361/cnnvd
- https://www.gehealthcare.com/en-US/securityPatchcnnvd
- https://us-cert.cisa.gov/ics/advisories/icsma-20-343-01enisa_euvd
- https://us-cert.cisa.gov/ics/advisories/icsma-20-343-01Third Party AdvisoryUS Government Resourcenvd