CVE-2018-7510

mediumpriority · track · 39EPSS 0.014 · 73 %ileEUUS
grimoire priorityP3 · Track
Exploitable?
—
no known exploitation
Likely / severe?
1.41% (73rd) EPSS
CVSS 5 · medium
Is it mine?
4 products
Check your exposure
Who says so?
4 sources
cnnvd · enisa_euvd · cisa_csaf_ot · nvd

EPSS · scored 2026-10-27

In the web application in BeaconMedaes TotalAlert Scroll Medical Air Systems running software versions prior to 4107600010.23, passwords are presented in plaintext in a file that is accessible without authentication.

01

remediations

Vendor fix, mitigation, or workaround guidance, as published by each source.

A vendor fix is available.

  • FixedTotalAlert Scroll Medical Air Systemscisa_csaf_ot

    BeaconMedaes has stated that the vulnerabilities do not compromise either patient health information or compliance with the NFPA 99 standard for healthcare facilities. To address these vulnerabilities, BeaconMedaes has created update 4107600010.24 and recommends that users of the TotalAlert Scroll Medical Air Systems update to this version or the latest release. BeaconMedaes recommends that affected users reach out to BeaconMedaes directly at 1-888-4MEDGAS (463-3427) to obtain this update.

02

cvss

Technical severity, 0–10. One row per scoring source.

9.8
010

base score 0–10 · higher = more severe · hover a vector to decode⚠ sources disagree · medium–critical · 5–9.8

v2.05mediumAV:N/AC:L/Au:N/C:P/I:N/A:Nprimary · NVD / NIST USexploitability 10 · impact 2.9
v3.09.8criticalCVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:Hprimary · NVD / NIST USexploitability 3.9 · impact 5.9
v3.07.5highCVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:Nsecondary · CISA US
v3.09criticalCVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H/E:Usecondary · CVSS-BT (t0sche)
03

weaknesses

The CWE weakness classes this flaw belongs to.

04

affected products (4)

Products and version ranges known to be affected.

05

advisories (1)

Vendor and ecosystem advisories describing this.

  • ICSMA-18-144-01BeaconMedaes TotalAlert Scroll Medical Air Systems2018-05-24icsa · cisa_csaf_otUS

    Successful exploitation of these vulnerabilities could allow an attacker to view and potentially modify some device information and web application setup information, which does not include access to patient health information. Additionally, BeaconMedaes has stated that a successful attacker would not be able to affect the ability of the device to operate as designed for the purpose of delivering medical air in compliance with the NFPA 99 standard for healthcare facilities.

06

also known as (3)

Other names and identifiers for this entity.

07

sources

Which sources assert this record, and when.

published 2018-06-06modified 2026-06-17
09

references (4)

External citations — patches, advisories, analysis.

scry

Attackers weaponise vulnerabilities like this against internet-exposed services. See what is exposed on your perimeter with a free passive scan.

Free scan