CVE-2018-18984
lowpriority · track · 18OT risk · elevatedEPSS 0.003 · 24 %ileEUUSEPSS · scored 2026-10-27
Medtronic CareLink and Encore Programmers do not encrypt or do not sufficiently encrypt sensitive PII and PHI information while at rest .
remediations
Vendor fix, mitigation, or workaround guidance, as published by each source.
A vendor fix is available.
- Fixed29901 Encore Programmercisa_csaf_ot
All affected programmers allow for the manual deletion of programmer-generated reports, which could contain PHI/PII. Medtronic recommends users delete these reports when no longer needed and prior to any disposition of the programmer.
- FixedCareLink 2090 Programmercisa_csaf_ot
All affected programmers allow for the manual deletion of programmer-generated reports, which could contain PHI/PII. Medtronic recommends users delete these reports when no longer needed and prior to any disposition of the programmer.
- FixedCareLink 9790 Programmercisa_csaf_ot
All affected programmers allow for the manual deletion of programmer-generated reports, which could contain PHI/PII. Medtronic recommends users delete these reports when no longer needed and prior to any disposition of the programmer.
cvss
Technical severity, 0–10. One row per scoring source.
base score 0–10 · higher = more severe · hover a vector to decode⚠ sources disagree · low–medium · 2.1–4.6
weaknesses
The CWE weakness classes this flaw belongs to.
affected products (12)
Products and version ranges known to be affected.
- medtronic 29901 encore programmer firmwareall versionsnvdUScpe:2.3:o:medtronic:29901_encore_programmer_firmware:*:*:*:*:*:*:*:*
- medtronic carelink 2090 programmer firmwareall versionsnvdUScpe:2.3:o:medtronic:carelink_2090_programmer_firmware:*:*:*:*:*:*:*:*
- medtronic carelink 9790 programmer firmwareall versionsnvdUScpe:2.3:o:medtronic:carelink_9790_programmer_firmware:*:*:*:*:*:*:*:*
advisories (1)
Vendor and ecosystem advisories describing this.
- ICSMA-18-347-01Medtronic 9790, 2090 CareLink, and 29901 Encore Programmers2018-12-13icsa · cisa_csaf_otUS
As part of the intended functionality of this device, it may store protected health information (PHI) or personally identifiable information (PII). Successful exploitation of the vulnerability may allow an attacker with physical access to an affected programmer to access PHI or PII stored on the device.
also known as (3)
Other names and identifiers for this entity.
sources
Which sources assert this record, and when.
references (9)
External citations — patches, advisories, analysis.
- https://global.medtronic.com/xg-en/product-security/security-bulletins/carelink-9790-2090-29901.htmlcnnvd
- https://ics-cert.us-cert.gov/advisories/ICSMA-18-347-01cnnvd
- https://www.securityfocus.com/bid/106215cnnvd
- http://www.securityfocus.com/bid/106215enisa_euvd
- https://global.medtronic.com/xg-en/product-security/security-bulletins/carelink-9790-2090-29901.htmlenisa_euvd
- https://ics-cert.us-cert.gov/advisories/ICSMA-18-347-01enisa_euvd
- http://www.securityfocus.com/bid/106215Third Party AdvisoryVDB Entrynvd