ICSMA-18-347-01

icsaUScisa_csaf_ot

Medtronic 9790, 2090 CareLink, and 29901 Encore Programmers

view advisory published 2018-12-13

As part of the intended functionality of this device, it may store protected health information (PHI) or personally identifiable information (PII). Successful exploitation of the vulnerability may allow an attacker with physical access to an affected programmer to access PHI or PII stored on the device.

01

describes (1)

Related CVEs.

IDseverity exploitable
CVE-2018-18984USmedium4.6OT elevated—
Medtronic CareLink and Encore Programmers do not encrypt or do not sufficiently encrypt sensitive PII and PHI information while at rest .
02

references

External citations — patches, advisories, analysis.

03

remediations (3)

  • Fixed29901 Encore Programmer

    All affected programmers allow for the manual deletion of programmer-generated reports, which could contain PHI/PII. Medtronic recommends users delete these reports when no longer needed and prior to any disposition of the programmer.

  • FixedCareLink 2090 Programmer

    All affected programmers allow for the manual deletion of programmer-generated reports, which could contain PHI/PII. Medtronic recommends users delete these reports when no longer needed and prior to any disposition of the programmer.

  • FixedCareLink 9790 Programmer

    All affected programmers allow for the manual deletion of programmer-generated reports, which could contain PHI/PII. Medtronic recommends users delete these reports when no longer needed and prior to any disposition of the programmer.