CVE-2017-12725
mediumpriority · track · 22OT risk · elevatedEPSS 0.008 · 57 %ileEUUSEPSS · scored 2026-10-27
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.
remediations
Vendor fix, mitigation, or workaround guidance, as published by each source.
A vendor fix is available.
- FixedMedfusion 4000 Wireless Syringe Infusion Pumpcisa_csaf_ot
Email: info.asd@smiths-medical.com
cvss
Technical severity, 0–10. One row per scoring source.
base score 0–10 · higher = more severe · hover a vector to decode⚠ sources disagree · low–medium · 3.7–6.8
weaknesses
The CWE weakness classes this flaw belongs to.
affected products (8)
Products and version ranges known to be affected.
- n/a Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump≤ Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pumpenisa_euvdEU
- smiths-medical medfusion 4000 wireless syringe infusion pump= 1.1nvdUScpe:2.3:o:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:*:*:*:*:*:*:*:*
- smiths-medical medfusion 4000 wireless syringe infusion pump= 1.5nvdUScpe:2.3:o:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:*:*:*:*:*:*:*:*
- smiths-medical medfusion 4000 wireless syringe infusion pump= 1.6nvdUScpe:2.3:o:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:*:*:*:*:*:*:*:*
advisories (1)
Vendor and ecosystem advisories describing this.
- ICSMA-17-250-02ASmiths Medical Medfusion 4000 Wireless Syringe Infusion Pump Vulnerabilities (Update A)2017-09-07icsa · cisa_csaf_otUS
This updated advisory is a follow-up to the original advisory titled ICSMA-17-250-02 Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump Vulnerabilities that was published September 7, 2017, on the NCCIC/ICS-CERT web site.
also known as (3)
Other names and identifiers for this entity.
sources
Which sources assert this record, and when.
references (7)
External citations — patches, advisories, analysis.
- http://www.securityfocus.com/bid/100665cnnvd
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02Acnnvd
- https://www.smiths-medical.com/Patchcnnvd
- http://www.securityfocus.com/bid/100665enisa_euvd
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02Aenisa_euvd
- http://www.securityfocus.com/bid/100665Third Party AdvisoryVDB Entrynvd
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02AThird Party AdvisoryUS Government Resourcenvd
Attackers weaponise vulnerabilities like this against internet-exposed services. See what is exposed on your perimeter with a free passive scan.