CVE-2017-12725

mediumpriority · track · 22OT risk · elevatedEPSS 0.008 · 57 %ileEUUS
grimoire priorityP3 · Track
Exploitable?
—
no known exploitation
Likely / severe?
0.83% (57th) EPSS
CVSS 6.8 · medium
Is it mine?
8 products
Check your exposure
Who says so?
4 sources
cnnvd · enisa_euvd · cisa_csaf_ot · nvd

EPSS · scored 2026-10-27

A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.

01

remediations

Vendor fix, mitigation, or workaround guidance, as published by each source.

A vendor fix is available.

  • FixedMedfusion 4000 Wireless Syringe Infusion Pumpcisa_csaf_ot

    Email: info.asd@smiths-medical.com

02

cvss

Technical severity, 0–10. One row per scoring source.

6.8
010

base score 0–10 · higher = more severe · hover a vector to decode⚠ sources disagree · low–medium · 3.7–6.8

v2.06.8mediumAV:N/AC:M/Au:N/C:P/I:P/A:Pprimary · NVD / NIST USexploitability 8.6 · impact 6.4
v3.05.6mediumCVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:Lprimary · NVD / NIST USexploitability 2.2 · impact 3.4
v3.03.7lowCVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:Nsecondary · CISA US
v3.05.1mediumCVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:L/E:Usecondary · CVSS-BT (t0sche)
03

weaknesses

The CWE weakness classes this flaw belongs to.

04

affected products (8)

Products and version ranges known to be affected.

05

advisories (1)

Vendor and ecosystem advisories describing this.

  • ICSMA-17-250-02ASmiths Medical Medfusion 4000 Wireless Syringe Infusion Pump Vulnerabilities (Update A)2017-09-07icsa · cisa_csaf_otUS

    This updated advisory is a follow-up to the original advisory titled ICSMA-17-250-02 Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump Vulnerabilities that was published September 7, 2017, on the NCCIC/ICS-CERT web site.

06

also known as (3)

Other names and identifiers for this entity.

07

sources

Which sources assert this record, and when.

published 2018-02-15modified 2026-06-17
09

references (7)

External citations — patches, advisories, analysis.

scry

Attackers weaponise vulnerabilities like this against internet-exposed services. See what is exposed on your perimeter with a free passive scan.