ICSMA-17-250-02A

icsaUScisa_csaf_ot

Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump Vulnerabilities (Update A)

view advisory published 2017-09-07

This updated advisory is a follow-up to the original advisory titled ICSMA-17-250-02 Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump Vulnerabilities that was published September 7, 2017, on the NCCIC/ICS-CERT web site.

01

describes (8)

Related CVEs.

IDseverity exploitable
CVE-2017-12726UShigh7.3OT elevated—
A Use of Hard-coded Password issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. Telnet on the pump uses hardcoded cred…
CVE-2017-12725USmedium5.6OT elevated—
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network confi…
CVE-2017-12724UShigh8.1OT elevated—
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP server on the pump contains…
CVE-2017-12723USlow3.7—
A Password in Configuration File issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump stores some passwords in…
CVE-2017-12722USmedium5.3—
An Out-of-bounds Read issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump read…
CVE-2017-12721USmedium5.9—
An Improper Certificate Validation issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump does not validate host…
CVE-2017-12720UShigh8.1OT elevated—
An Improper Access Control issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP server on the pump does not requ…
CVE-2017-12718UShigh8.1Exploit
A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump…
02

references

External citations — patches, advisories, analysis.

03

remediations (8)

  • FixedMedfusion 4000 Wireless Syringe Infusion Pump

    Email: info.asd@smiths-medical.com

  • FixedMedfusion 4000 Wireless Syringe Infusion Pump

    Email: info.asd@smiths-medical.com

  • FixedMedfusion 4000 Wireless Syringe Infusion Pump

    Email: info.asd@smiths-medical.com

  • FixedMedfusion 4000 Wireless Syringe Infusion Pump

    Email: info.asd@smiths-medical.com

  • FixedMedfusion 4000 Wireless Syringe Infusion Pump

    Email: info.asd@smiths-medical.com

  • FixedMedfusion 4000 Wireless Syringe Infusion Pump

    Email: info.asd@smiths-medical.com

  • FixedMedfusion 4000 Wireless Syringe Infusion Pump

    Email: info.asd@smiths-medical.com

  • FixedMedfusion 4000 Wireless Syringe Infusion Pump

    Email: info.asd@smiths-medical.com