CVE-2014-5433

highpriority · track · 40OT risk · elevatedEPSS 0.021 · 81 %ileEUUS
grimoire priorityP2 · Scheduled
Exploitable?
—
no known exploitation
Likely / severe?
2.06% (81st) EPSS
CVSS 7.5 · high
Is it mine?
2 products
Check your exposure
Who says so?
3 sources
cnnvd · enisa_euvd · nvd

EPSS · scored 2026-10-28

An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16, which may allow an attacker to gain access the host network. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.

01

cvss

Technical severity, 0–10. One row per scoring source.

9.8
010

base score 0–10 · higher = more severe · hover a vector to decode⚠ sources disagree · high–critical · 7.5–9.8

v2.07.5highAV:N/AC:L/Au:N/C:P/I:P/A:Pprimary · NVD / NIST USexploitability 10 · impact 6.4
v3.09.8criticalCVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:Hprimary · NVD / NIST USexploitability 3.9 · impact 5.9
v3.09criticalCVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H/E:Usecondary · CVSS-BT (t0sche)
02

weaknesses

The CWE weakness classes this flaw belongs to.

03

affected products (2)

Products and version ranges known to be affected.

04

also known as (3)

Other names and identifiers for this entity.

05

sources

Which sources assert this record, and when.

published 2019-03-26modified 2026-06-17
07

references (5)

External citations — patches, advisories, analysis.

scry

Attackers weaponise vulnerabilities like this against internet-exposed services. See what is exposed on your perimeter with a free passive scan.

Free scan