CVE-2018-10626

lowPriorität · verfolgen · 18OT-Risiko · erhöhtEPSS 0.004 · 30. PerzentilEURUUS
grimoire priorityP3 · Track
Exploitable?
—
no known exploitation
Likely / severe?
0.36% (30th) EPSS
CVSS 3.8 · low
Is it mine?
10 products
Prüfe deine Angriffsfläche
Who says so?
5 sources
cnnvd · enisa_euvd · fstec_bdu · cisa_csaf_ot

EPSS · bewertet 2026-10-27

Medtronic MyCareLink Patient Monitor’s update service does not sufficiently verify the authenticity of the data uploaded. An attacker who obtains per-product credentials from the monitor and paired implantable cardiac device information can potentially upload invalid data to the Medtronic CareLink network.

01

Behebungen

Hersteller-Fix, Mitigation oder Workaround, wie von den Quellen veröffentlicht.

Noch kein Hersteller-Fix; eine Mitigation ist verfügbar.

  • Mitigation24950 MyCareLink Monitorcisa_csaf_ot

    Maintain good physical control over the home monitor.

  • Mitigation24952 MyCareLink Monitorcisa_csaf_ot

    Maintain good physical control over the home monitor.

02

cvss

Technischer Schweregrad, 0–10. Eine Zeile je Quelle.

4.4
010

Basiswert 0–10 · höher = schwerwiegender · Vektor überfahren zum Auflösen⚠ Quellen uneinig · low–medium · 3.8–4.4

v2.03.8lowAV:A/AC:M/Au:S/C:P/I:P/A:Nprimary · NVD / NIST USAusnutzbarkeit 4.4 · Auswirkung 4.9
v3.04.4mediumCVSS:3.0/AV:A/AC:H/PR:L/UI:N/S:C/C:L/I:L/A:Nprimary · NVD / NIST USAusnutzbarkeit 1.3 · Auswirkung 2.7
v3.14.4mediumCVSS:3.1/AV:A/AC:H/PR:L/UI:N/S:C/C:L/I:L/A:Nsecondary · CISA US
v3.14.4mediumAV:A/AC:H/PR:L/UI:N/S:C/C:L/I:L/A:Nsecondary · FSTEC (BDU) RU
v3.14.4mediumCVSS:3.1/AV:A/AC:H/PR:L/UI:N/S:C/C:L/I:L/A:Nsecondary · ENISA EU
v3.14.4mediumCVSS:3.1/AV:A/AC:H/PR:L/UI:N/S:C/C:L/I:L/A:Nsecondary · ics-cert@hq.dhs.gov USAusnutzbarkeit 1.3 · Auswirkung 2.7
v3.14.1mediumCVSS:3.1/AV:A/AC:H/PR:L/UI:N/S:C/C:L/I:L/A:N/E:Usecondary · CVSS-BT (t0sche)
03

schwächen

Die CWE-Schwächeklassen, zu denen dies gehört.

04

betroffene Produkte (10)

Betroffene Produkte und Versionsbereiche.

05

Advisories (1)

Hersteller- und Ökosystem-Meldungen dazu.

  • ICSMA-18-219-01Medtronic MyCareLink 24950 Patient Monitor (Update A)2018-08-07icsa · cisa_csaf_otUS

    Successful exploitation of these vulnerabilities may allow an attacker with physical access to obtain per-product credentials that are utilized to authenticate data uploads and encrypt data at rest. Additionally, an attacker with access to a set of these credentials and additional identifiers can upload invalid data to the Medtronic CareLink network.

06

auch bekannt als (4)

Weitere Namen und Kennungen dieser Entität.

07

quellen

Welche Quellen diesen Datensatz belegen und wann.

veröffentlicht 2018-08-10geändert 2026-06-17
09

Referenzen (15)

Externe Belege — Patches, Meldungen, Analysen.

scry

Angreifer nutzen Schwachstellen wie diese gegen im Internet exponierte Dienste aus. Sehen Sie mit einem kostenlosen Passiv-Scan, was exponiert ist.

Kostenloser Scan