Medtronic MyCareLink 24950 Patient Monitor (Update A)
Successful exploitation of these vulnerabilities may allow an attacker with physical access to obtain per-product credentials that are utilized to authenticate data uploads and encrypt data at rest. Additionally, an attacker with access to a set of these credentials and additional identifiers can upload invalid data to the Medtronic CareLink network.
beschreibt (2)
Zugehörige CVEs.
| ID | Schweregrad | ausnutzbar | ||
|---|---|---|---|---|
| CVE-2018-10626US | medium4.4OT erhöht | — | ||
Medtronic MyCareLink Patient Monitor’s update service does not sufficiently verify the authenticity of the data uploaded. An
attacker who obtains per-product credentials from the… | ||||
| CVE-2018-10622US | high7.1 | — | ||
Medtronic MyCareLink Patient Monitor uses per-product credentials that are stored in a recoverable format. An attacker can use these credentials for network authentication. | ||||
Referenzen
Externe Belege — Patches, Meldungen, Analysen.
- ICS Advisory ICSMA-18-219-01 JSONself
- Recommended Practicesexternal
- ICSA Advisory ICSMA-18-219-01 - Web Versionself
- Recommended Practicesexternal
- Recommended Practicesexternal
- Recommended Practicesexternal
- Recommended Practicesexternal
- Recommended Practicesexternal
Behebungen (4)
- Mitigation24950 MyCareLink Monitor
Maintain good physical control over the home monitor.
- Mitigation24950 MyCareLink Monitor
Maintain good physical control over the home monitor.
- Mitigation24952 MyCareLink Monitor
Maintain good physical control over the home monitor.
- Mitigation24952 MyCareLink Monitor
Maintain good physical control over the home monitor.