Hospira LifeCare PCA Infusion System Vulnerabilities
This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov
beschreibt (7)
Zugehörige CVEs.
| ID | Schweregrad | ausnutzbar | ||
|---|---|---|---|---|
| CVE-2015-1012US | high7.5OT erhöht | — | ||
Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicate… | ||||
| CVE-2015-3958US | high7.8 | — | ||
Hospira LifeCare PCA Infusion System 5.0 and earlier, and possibly other versions, allows remote attackers to cause a denial of service (forced manual reboot) via a flood of TCP pa… | ||||
| CVE-2015-3957US | medium4.6OT erhöht | — | ||
Hospira LifeCare PCA Infusion System before 7.0 stores private keys and certificates, which has unspecified impact and attack vectors. | ||||
| CVE-2015-3955US | high10.0 | — | ||
Stack-based buffer overflow in Hospira LifeCare PCA Infusion System 5.0 and earlier, and possibly other versions, allows remote attackers to execute arbitrary code via unspecified… | ||||
| CVE-2015-1011US | medium5.0OT erhöht | — | ||
Hospira LifeCare PCA Infusion System before 7.0 has hardcoded credentials, which makes it easier for remote attackers to obtain access via unspecified vectors. | ||||
| CVE-2014-5406US | high9.3OT erhöht | — | ||
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change,… | ||||
| CVE-2015-3459US | high10.0 | — | ||
The communication module on the Hospira LifeCare PCA Infusion System before 7.0 does not require authentication for root TELNET sessions, which allows remote attackers to modify th… | ||||
Referenzen
Externe Belege — Patches, Meldungen, Analysen.
- ICS Advisory ICSA-15-125-01B JSONself
- Recommended Practicesexternal
- ICS Advisory ICSA-15-125-01B - Web Versionself
- Recommended Practicesexternal
- Recommended Practicesexternal
- Recommended Practicesexternal
- Recommended Practicesexternal
- Recommended Practicesexternal
Behebungen (7)
- MitigationLifeCare PCA Infusion System
Existing PCA Infusion Systems running Version 5.0 can be upgraded to Version 7.0 when it becomes available. Hospira will be retiring older versions of the LifeCare PCA Infusion System, Versions 2 and Versions 3, by the end of the year, 2015.
- MitigationLifeCare PCA Infusion System
Existing PCA Infusion Systems running Version 5.0 can be upgraded to Version 7.0 when it becomes available. Hospira will be retiring older versions of the LifeCare PCA Infusion System, Versions 2 and Versions 3, by the end of the year, 2015.
- MitigationLifeCare PCA Infusion System
Existing PCA Infusion Systems running Version 5.0 can be upgraded to Version 7.0 when it becomes available. Hospira will be retiring older versions of the LifeCare PCA Infusion System, Versions 2 and Versions 3, by the end of the year, 2015.
- MitigationLifeCare PCA Infusion System
Existing PCA Infusion Systems running Version 5.0 can be upgraded to Version 7.0 when it becomes available. Hospira will be retiring older versions of the LifeCare PCA Infusion System, Versions 2 and Versions 3, by the end of the year, 2015.
- MitigationLifeCare PCA Infusion System
Existing PCA Infusion Systems running Version 5.0 can be upgraded to Version 7.0 when it becomes available. Hospira will be retiring older versions of the LifeCare PCA Infusion System, Versions 2 and Versions 3, by the end of the year, 2015.
- MitigationLifeCare PCA Infusion System
Existing PCA Infusion Systems running Version 5.0 can be upgraded to Version 7.0 when it becomes available. Hospira will be retiring older versions of the LifeCare PCA Infusion System, Versions 2 and Versions 3, by the end of the year, 2015.
- MitigationLifeCare PCA Infusion System
Existing PCA Infusion Systems running Version 5.0 can be upgraded to Version 7.0 when it becomes available. Hospira will be retiring older versions of the LifeCare PCA Infusion System, Versions 2 and Versions 3, by the end of the year, 2015.
Advisories wie diese benennen die exponierten Produkte, die Angreifer auf Ihrer Angriffsfläche ins Visier nehmen. Sehen Sie mit einem kostenlosen Passiv-Scan, was exponiert ist.
Kostenloser Scan