CVE-2014-5406
highPriorität · verfolgen · 37OT-Risiko · erhöhtEPSS 0.012 · 68. PerzentilEUUSEPSS · bewertet 2026-10-22
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.
Behebungen
Hersteller-Fix, Mitigation oder Workaround, wie von den Quellen veröffentlicht.
Noch kein Hersteller-Fix; eine Mitigation ist verfügbar.
- MitigationLifeCare PCA Infusion Systemcisa_csaf_ot
Existing PCA Infusion Systems running Version 5.0 can be upgraded to Version 7.0 when it becomes available. Hospira will be retiring older versions of the LifeCare PCA Infusion System, Versions 2 and Versions 3, by the end of the year, 2015.
cvss
Technischer Schweregrad, 0–10. Eine Zeile je Quelle.
Basiswert 0–10 · höher = schwerwiegender · Vektor überfahren zum Auflösen
schwächen
Die CWE-Schwächeklassen, zu denen dies gehört.
betroffene Produkte (4)
Betroffene Produkte und Versionsbereiche.
- hospira lifecare pcainfusion firmware≤ 5.0nvdUScpe:2.3:o:hospira:lifecare_pcainfusion_firmware:*:*:*:*:*:*:*:*
Advisories (1)
Hersteller- und Ökosystem-Meldungen dazu.
- ICSA-15-125-01BHospira LifeCare PCA Infusion System Vulnerabilities2015-02-05icsa · cisa_csaf_otUS
This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov
auch bekannt als (3)
Weitere Namen und Kennungen dieser Entität.
quellen
Welche Quellen diesen Datensatz belegen und wann.
Referenzen (14)
Externe Belege — Patches, Meldungen, Analysen.
- http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htmcnnvd
- http://www.hospira.com/Patchcnnvd
- http://www.securityfocus.com/bid/74476cnnvd
- https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01cnnvd
- https://xs-sniper.com/blog/2015/06/08/hospira-plum-a-infusion-pump-vulnerabilities/cnnvd
- http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htmenisa_euvd
- https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2015/icsa-15-125-01.jsonenisa_euvd
- https://www.cisa.gov/news-events/ics-advisories/icsa-15-125-01enisa_euvd
Angreifer nutzen Schwachstellen wie diese gegen im Internet exponierte Dienste aus. Sehen Sie mit einem kostenlosen Passiv-Scan, was exponiert ist.
Kostenloser Scan