ICSMA-23-061-01

icsaUScisa_csaf_ot

Medtronic Micro Clinician and InterStim Apps

view advisory published 2023-03-06

Successful exploitation of this vulnerability could cause the clinician application’s custom password to be reset to default, resulting in unauthorized control of the clinician therapy application.

01

describes (1)

Related CVEs.

IDseverity exploitable
CVE-2023-25931USmedium6.8—
Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update…
02

references

External citations — patches, advisories, analysis.

03

remediations (2)

  • FixedInterStim X Clinician

    An app update is available as of February 23, 2023 that will fix the vulnerability.

  • FixedMicro Clinician

    An app update is available as of February 23, 2023 that will fix the vulnerability.