CVE-2023-25931

mediumpriority · track · 27EPSS 0.003 · 19 %ileEUUS
grimoire priorityP3 · Track
Exploitable?
—
no known exploitation
Likely / severe?
0.27% (19th) EPSS
CVSS 6.8 · medium
Is it mine?
8 products
Check your exposure
Who says so?
4 sources
cnnvd · enisa_euvd · cisa_csaf_ot · nvd

EPSS · scored 2026-10-27

Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.

01

remediations

Vendor fix, mitigation, or workaround guidance, as published by each source.

A vendor fix is available.

  • FixedInterStim X Cliniciancisa_csaf_ot

    An app update is available as of February 23, 2023 that will fix the vulnerability.

  • FixedMicro Cliniciancisa_csaf_ot

    An app update is available as of February 23, 2023 that will fix the vulnerability.

02

cvss

Technical severity, 0–10. One row per scoring source.

6.8
010

base score 0–10 · higher = more severe · hover a vector to decode

v3.16.8mediumCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:Hprimary · NVD / NIST USexploitability 0.9 · impact 5.9
v3.06.4mediumCVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:Hsecondary · CISA US
v3.16.4mediumCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:Hsecondary · ENISA EU
v3.16.4mediumCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:Hsecondary · security@medtronic.comexploitability 0.9 · impact 5.5
v3.16.2mediumCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H/E:Usecondary · CVSS-BT (t0sche)
03

ssvc

A stakeholder decision — act, attend, or track — from exploitation and impact.

Trackcoordinatorcisa_ssvcUS
04

weaknesses

The CWE weakness classes this flaw belongs to.

05

affected products (8)

Products and version ranges known to be affected.

06

advisories (1)

Vendor and ecosystem advisories describing this.

  • ICSMA-23-061-01Medtronic Micro Clinician and InterStim Apps2023-03-06icsa · cisa_csaf_otUS

    Successful exploitation of this vulnerability could cause the clinician application’s custom password to be reset to default, resulting in unauthorized control of the clinician therapy application.

07

also known as (3)

Other names and identifiers for this entity.

08

sources

Which sources assert this record, and when.

published 2023-03-01modified 2026-06-17
10

references (5)

External citations — patches, advisories, analysis.

scry

Attackers weaponise vulnerabilities like this against internet-exposed services. See what is exposed on your perimeter with a free passive scan.

Free scan