CVE-2024-1275
highpriority · track · 0EPSS 0.004 · 33 %ileEUUSEPSS · scored 2026-10-27
Use of Default Cryptographic Key vulnerability in Baxter Welch Allyn Connex Spot Monitor may allow Configuration/Environment Manipulation.This issue affects Welch Allyn Connex Spot Monitor in all versions prior to 1.52.
remediations
Vendor fix, mitigation, or workaround guidance, as published by each source.
No vendor fix yet; a mitigation is available.
- MitigationWelch Allyn Connex Spot Monitor (CSM)cisa_csaf_ot
Apply proper network and physical security controls.
cvss
Technical severity, 0–10. One row per scoring source.
base score 0–10 · higher = more severe · hover a vector to decode⚠ sources disagree · medium–critical · 6.9–9.1
ssvc
A stakeholder decision — act, attend, or track — from exploitation and impact.
weaknesses
The CWE weakness classes this flaw belongs to.
affected products (3)
Products and version ranges known to be affected.
advisories (1)
Vendor and ecosystem advisories describing this.
- ICSMA-24-151-02Baxter Welch Allyn Connex Spot Monitor2024-05-30icsa · cisa_csaf_otUS
Successful exploitation of this vulnerability could allow an attacker to modify device configuration and firmware data. Tampering with this data could lead to device compromise, resulting in impact and/or delay in patient care.
also known as (3)
Other names and identifiers for this entity.
sources
Which sources assert this record, and when.
references (5)
External citations — patches, advisories, analysis.
- https://cxsecurity.com/cveshow/CVE-2024-1275/cnnvd
- https://www.baxter.com/sites/g/files/ebysai3896/files/2024-05/80030693A_ICMSA.pdfPatchcnnvd
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-151-02cnnvd
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-151-02enisa_euvd
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-151-02nvd