CVE-2022-32537

mediumpriority · track · 19EPSS 0.003 · 28 %ileEUUS
grimoire priorityP3 · Track
Exploitable?
—
no known exploitation
Likely / severe?
0.34% (28th) EPSS
CVSS 4.8 · medium
Is it mine?
37 products
Check your exposure
Who says so?
4 sources
cnnvd · enisa_euvd · cisa_csaf_ot · nvd

EPSS · scored 2026-10-27

A vulnerability exists which could allow an unauthorized user to learn aspects of the communication protocol used to pair system components while the pump is being paired with other system components. Exploitation requires nearby wireless signal proximity with the patient and the device; advanced technical knowledge is required for exploitation. Please refer to the Medtronic Product Security Bulletin for guidance

01

remediations

Vendor fix, mitigation, or workaround guidance, as published by each source.

A vendor fix is available.

  • FixedMiniMed 620Gcisa_csaf_ot

    Medtronic has identified the following precautions to assist users:

  • FixedMiniMed 630Gcisa_csaf_ot

    Medtronic has identified the following precautions to assist users:

  • FixedMiniMed 640Gcisa_csaf_ot

    Medtronic has identified the following precautions to assist users:

  • FixedMiniMed 670Gcisa_csaf_ot

    Medtronic has identified the following precautions to assist users:

02

cvss

Technical severity, 0–10. One row per scoring source.

4.8
010

base score 0–10 · higher = more severe · hover a vector to decode

v3.14.8mediumCVSS:3.1/AV:A/AC:H/PR:L/UI:N/S:U/C:N/I:H/A:Nprimary · NVD / NIST USexploitability 1.2 · impact 3.6
v3.04.8mediumCVSS:3.0/AV:A/AC:H/PR:L/UI:N/S:U/C:N/I:H/A:Nsecondary · CISA US
v3.14.8mediumCVSS:3.1/AV:A/AC:H/PR:L/UI:N/S:U/C:N/I:H/A:Nsecondary · ENISA EU
v3.14.8mediumCVSS:3.1/AV:A/AC:H/PR:L/UI:N/S:U/C:N/I:H/A:Nsecondary · security@medtronic.comexploitability 1.2 · impact 3.6
v3.14.4mediumCVSS:3.1/AV:A/AC:H/PR:L/UI:N/S:U/C:N/I:H/A:N/E:Usecondary · CVSS-BT (t0sche)
03

ssvc

A stakeholder decision — act, attend, or track — from exploitation and impact.

Trackcoordinatorcisa_ssvcUS
04

weaknesses

The CWE weakness classes this flaw belongs to.

05

affected products (37)

Products and version ranges known to be affected.

showing 20 of 37
06

advisories (1)

Vendor and ecosystem advisories describing this.

  • ICSMA-22-263-01Medtronic NGP 600 Series Insulin Pumps2022-09-20icsa · cisa_csaf_otUS

    Successful exploitation of this vulnerability could allow an unauthorized user to deliver too much or too little insulin through delivery of an unintended insulin bolus or because insulin delivery is slowed or stopped.

07

also known as (3)

Other names and identifiers for this entity.

08

sources

Which sources assert this record, and when.

published 2022-12-12modified 2026-06-17
09

references (12)

External citations — patches, advisories, analysis.

scry

Attackers weaponise vulnerabilities like this against internet-exposed services. See what is exposed on your perimeter with a free passive scan.

Free scan