CVE-2022-26393

highpriority · track · 32EPSS 0.007 · 52 %ileEUUS
grimoire priorityP2 · Scheduled
Exploitable?
—
no known exploitation
Likely / severe?
0.67% (52nd) EPSS
CVSS 8.1 · high
Is it mine?
16 products
Check your exposure
Who says so?
4 sources
cnnvd · enisa_euvd · cisa_csaf_ot · nvd

EPSS · scored 2026-10-27

The Baxter Spectrum WBM is susceptible to format string attacks via application messaging. An attacker could use this to read memory in the WBM to access sensitive information or cause a Denial of Service (DoS) on the WBM.

01

remediations

Vendor fix, mitigation, or workaround guidance, as published by each source.

A vendor fix is available.

  • FixedBaxter Spectrum IQcisa_csaf_ot

    According to Baxter, software updates to disable Telnet and FTP (CVE-2022-26392) are in process. Software updates addressing the format string attack (CVE-2022-26393) are included in WBM version 20D30 and all other WBM versions authentication is already available in Spectrum IQ (CVE-2022-26394). Instructions to erase all data and settings on WBMs and pumps before decommissioning and transferring to other facilities (CVE-2022-26390) are in process for incorporation into the Spectrum Operator's Manual.

  • FixedBaxter Spectrum IQ LVPcisa_csaf_ot

    According to Baxter, software updates to disable Telnet and FTP (CVE-2022-26392) are in process. Software updates addressing the format string attack (CVE-2022-26393) are included in WBM version 20D30 and all other WBM versions authentication is already available in Spectrum IQ (CVE-2022-26394). Instructions to erase all data and settings on WBMs and pumps before decommissioning and transferring to other facilities (CVE-2022-26390) are in process for incorporation into the Spectrum Operator's Manual.

  • FixedSigma Spectrumcisa_csaf_ot

    According to Baxter, software updates to disable Telnet and FTP (CVE-2022-26392) are in process. Software updates addressing the format string attack (CVE-2022-26393) are included in WBM version 20D30 and all other WBM versions authentication is already available in Spectrum IQ (CVE-2022-26394). Instructions to erase all data and settings on WBMs and pumps before decommissioning and transferring to other facilities (CVE-2022-26390) are in process for incorporation into the Spectrum Operator's Manual.

  • FixedSigma Spectrum LVPcisa_csaf_ot

    According to Baxter, software updates to disable Telnet and FTP (CVE-2022-26392) are in process. Software updates addressing the format string attack (CVE-2022-26393) are included in WBM version 20D30 and all other WBM versions authentication is already available in Spectrum IQ (CVE-2022-26394). Instructions to erase all data and settings on WBMs and pumps before decommissioning and transferring to other facilities (CVE-2022-26390) are in process for incorporation into the Spectrum Operator's Manual.

  • MitigationBaxter Spectrum IQcisa_csaf_ot

    As a last resort, users may disable wireless operation of the pump; the Spectrum Infusion System was designed to operate without network access. This action would impact an organization’s ability to rapidly deploy drug library (formulary) updates to their pumps.

  • MitigationBaxter Spectrum IQ LVPcisa_csaf_ot

    As a last resort, users may disable wireless operation of the pump; the Spectrum Infusion System was designed to operate without network access. This action would impact an organization’s ability to rapidly deploy drug library (formulary) updates to their pumps.

  • MitigationSigma Spectrumcisa_csaf_ot

    As a last resort, users may disable wireless operation of the pump; the Spectrum Infusion System was designed to operate without network access. This action would impact an organization’s ability to rapidly deploy drug library (formulary) updates to their pumps.

  • MitigationSigma Spectrum LVPcisa_csaf_ot

    As a last resort, users may disable wireless operation of the pump; the Spectrum Infusion System was designed to operate without network access. This action would impact an organization’s ability to rapidly deploy drug library (formulary) updates to their pumps.

02

cvss

Technical severity, 0–10. One row per scoring source.

8.1
010

base score 0–10 · higher = more severe · hover a vector to decode⚠ sources disagree · medium–high · 5–8.1

v3.18.1highCVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:H/I:N/A:Hprimary · NVD / NIST USexploitability 2.8 · impact 5.2
v3.05mediumCVSS:3.0/AV:N/AC:H/PR:L/UI:N/S:U/C:L/I:L/A:Lsecondary · CISA US
v3.15mediumCVSS:3.1/AV:N/AC:H/PR:L/UI:N/S:U/C:L/I:L/A:Lsecondary · ENISA EU
v3.15mediumCVSS:3.1/AV:N/AC:H/PR:L/UI:N/S:U/C:L/I:L/A:Lsecondary · productsecurity@baxter.comexploitability 1.6 · impact 3.4
v3.17.4highCVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:H/I:N/A:H/E:Usecondary · CVSS-BT (t0sche)
03

weaknesses

The CWE weakness classes this flaw belongs to.

04

affected products (16)

Products and version ranges known to be affected.

05

advisories (1)

Vendor and ecosystem advisories describing this.

  • ICSMA-22-251-01Baxter Sigma Spectrum Infusion Pump (Update A)2022-09-08icsa · cisa_csaf_otUS

    Successful exploitation of these vulnerabilities could result in access to sensitive data and alteration of system configuration.

06

also known as (3)

Other names and identifiers for this entity.

07

sources

Which sources assert this record, and when.

published 2022-09-09modified 2026-06-17
09

references (8)

External citations — patches, advisories, analysis.

scry

Attackers weaponise vulnerabilities like this against internet-exposed services. See what is exposed on your perimeter with a free passive scan.

Free scan