CVE-2020-25166
highpriority · track · 28EPSS 0.005 · 40 %ileEUUSEPSS · scored 2026-10-23
An improper verification of the cryptographic signature of firmware updates of the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers to generate valid firmware updates with arbitrary content that can be used to tamper with devices.
remediations
Vendor fix, mitigation, or workaround guidance, as published by each source.
A vendor fix is available.
- FixedBattery pack with Wi-Ficisa_csaf_ot
Users in the United States and Canada who need additional support can contact B. Braun Technical Support by calling 800-627-PUMP or by emailing AISTechSupport@bbraunusa.com.
- FixedData module compactpluscisa_csaf_ot
Users in the United States and Canada who need additional support can contact B. Braun Technical Support by calling 800-627-PUMP or by emailing AISTechSupport@bbraunusa.com.
- FixedSpaceComcisa_csaf_ot
Users in the United States and Canada who need additional support can contact B. Braun Technical Support by calling 800-627-PUMP or by emailing AISTechSupport@bbraunusa.com.
- MitigationBattery pack with Wi-Ficisa_csaf_ot
All facilities utilizing SpaceCom 2, and Battery Pack SP with Wi-Fi should review their IT infrastructure to ensure a network zone concept has been implemented whereby critical systems, such as infusion pumps, are housed in separate (e.g., by firewalls or VLAN) environments that are not accessible directly from the Internet or by unauthorized users.
- MitigationData module compactpluscisa_csaf_ot
All facilities utilizing SpaceCom 2, and Battery Pack SP with Wi-Fi should review their IT infrastructure to ensure a network zone concept has been implemented whereby critical systems, such as infusion pumps, are housed in separate (e.g., by firewalls or VLAN) environments that are not accessible directly from the Internet or by unauthorized users.
- MitigationSpaceComcisa_csaf_ot
All facilities utilizing SpaceCom 2, and Battery Pack SP with Wi-Fi should review their IT infrastructure to ensure a network zone concept has been implemented whereby critical systems, such as infusion pumps, are housed in separate (e.g., by firewalls or VLAN) environments that are not accessible directly from the Internet or by unauthorized users.
cvss
Technical severity, 0–10. One row per scoring source.
base score 0–10 · higher = more severe · hover a vector to decode⚠ sources disagree · medium–high · 6.5–7.6
ssvc
A stakeholder decision — act, attend, or track — from exploitation and impact.
weaknesses
The CWE weakness classes this flaw belongs to.
affected products (15)
Products and version ranges known to be affected.
advisories (1)
Vendor and ecosystem advisories describing this.
- ICSMA-20-296-02B. Braun SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus2020-10-22icsa · cisa_csaf_otUS
Successful exploitation of these vulnerabilities could allow an attacker to compromise the security of the Space or compactplus communication devices, allowing an attacker to escalate privileges, view sensitive information, upload arbitrary files, and perform remote code execution.
also known as (3)
Other names and identifiers for this entity.
sources
Which sources assert this record, and when.
references (10)
External citations — patches, advisories, analysis.
- https://cxsecurity.com/cveshow/CVE-2020-25166/cnnvd
- https://us-cert.cisa.gov/ics/advisories/icsma-20-296-02cnnvd
- https://www.auscert.org.au/bulletins/ESB-2020.3661/cnnvd
- https://www.auscert.org.au/bulletins/ESB-2022.5282cnnvd
- https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.htmlcnnvd
- https://www.cisa.gov/uscert/ics/advisories/icsma-20-296-02cnnvd
- https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.htmlenisa_euvd
- https://www.cisa.gov/uscert/ics/advisories/icsma-20-296-02enisa_euvd
Attackers weaponise vulnerabilities like this against internet-exposed services. See what is exposed on your perimeter with a free passive scan.