Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35700BAX2 contain hardcoded passwords when physically entered on the keypad provide access to biomedical menus including device settings, view calibration values, network configuration of Sigma Spectrum WBM if installed.

01

remediations

Vendor fix, mitigation, or workaround guidance, as published by each source.

A vendor fix is available.

  • FixedBaxter Spectrumcisa_csaf_ot

    As a last resort, users may disable wireless operation of the pump. The Spectrum Infusion System was designed to operate without network access. This action would impact an organization 's ability to rapidly deploy drug library (formulary) updates to their pumps.

  • FixedBaxter Spectrum LVPcisa_csaf_ot

    As a last resort, users may disable wireless operation of the pump. The Spectrum Infusion System was designed to operate without network access. This action would impact an organization 's ability to rapidly deploy drug library (formulary) updates to their pumps.

  • FixedSigma Spectrumcisa_csaf_ot

    As a last resort, users may disable wireless operation of the pump. The Spectrum Infusion System was designed to operate without network access. This action would impact an organization 's ability to rapidly deploy drug library (formulary) updates to their pumps.

  • FixedSigma Spectrum LVPcisa_csaf_ot

    As a last resort, users may disable wireless operation of the pump. The Spectrum Infusion System was designed to operate without network access. This action would impact an organization 's ability to rapidly deploy drug library (formulary) updates to their pumps.

02

cvss

Technical severity, 0–10. One row per scoring source.

4.3
010

base score 0–10 · higher = more severe · hover a vector to decode⚠ sources disagree · low–medium · 2.1–4.3

v2.02.1lowAV:L/AC:L/Au:N/C:P/I:N/A:Nprimary · NVD / NIST USexploitability 3.9 · impact 2.9
v3.12.4lowCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:N/A:Nprimary · NVD / NIST USexploitability 0.9 · impact 1.4
v3.04.3mediumCVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:Lsecondary · CISA US
v3.12.2lowCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:N/A:N/E:Usecondary · CVSS-BT (t0sche)
03

weaknesses

The CWE weakness classes this flaw belongs to.

04

affected products (14)

Products and version ranges known to be affected.

05

advisories (1)

Vendor and ecosystem advisories describing this.

  • ICSMA-20-170-04Baxter Sigma Spectrum Infusion Pumps (Update B)2020-06-18icsa · cisa_csaf_otUS

    Successful exploitation of these vulnerabilities could result in access to sensitive data, alteration of system configuration, and impact to system availability.

06

also known as (3)

Other names and identifiers for this entity.

07

sources

Which sources assert this record, and when.

published 2020-06-29modified 2026-06-17
09

references (6)

External citations — patches, advisories, analysis.

scry

Attackers weaponise vulnerabilities like this against internet-exposed services. See what is exposed on your perimeter with a free passive scan.

Free scan