CVE-2017-9657
lowpriority · track · 26EPSS 0.008 · 55 %ileEUUSEPSS · scored 2026-10-22
Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible. In this state, the central monitoring station can indicate the MX40 is not connected or associated to the central monitor, and thus should be operating in local monitoring mode (local audio-on, screen-on), but the MX40 WLAN itself can instead still be operating in telemetry mode (local audio-off, screen-off). If a patient experiences an alarm event and clinical staff expects the MX40 to provide local alarming when it is not available from the local device, a delay of treatment can occur. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.
remediations
Vendor fix, mitigation, or workaround guidance, as published by each source.
No vendor fix yet; a mitigation is available.
- MitigationIntelliVue MX40 Patient Worn Monitor (WLAN only)cisa_csaf_ot
Philips is planning to release an additional MX40 software update in 2017 to address the improper handling of exceptional conditions vulnerability. Please see the Philips product security web site for latest information for this and other Philips products:
https://www.usa.philips.com/healthcare/about/customer-support/product-security
cvss
Technical severity, 0–10. One row per scoring source.
base score 0–10 · higher = more severe · hover a vector to decode⚠ sources disagree · low–medium · 3.3–6.5
weaknesses
The CWE weakness classes this flaw belongs to.
affected products (4)
Products and version ranges known to be affected.
- Philips IntelliVue MX40 Patient Worn Monitor≤ IntelliVue MX40 Patient Worn Monitor (WLAN only), all versions prior to Version B.06.18enisa_euvdEU
- philips intellivue mx40 firmware< b.06.18nvdUScpe:2.3:o:philips:intellivue_mx40_firmware:*:*:*:*:*:*:*:*
advisories (1)
Vendor and ecosystem advisories describing this.
- ICSMA-17-255-01ICSMA-17-255-01_Philips' IntelliView MX40 Patient Worn Monitor (WLAN) Vulnerabilities2017-09-12icsa · cisa_csaf_otUS
This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov
also known as (3)
Other names and identifiers for this entity.
sources
Which sources assert this record, and when.
references (9)
External citations — patches, advisories, analysis.
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-255-01cnnvd
- https://www.securityfocus.com/bid/100813cnnvd
- https://www.usa.philips.com/healthcare/about/customer-support/product-securitycnnvd
- http://www.securityfocus.com/bid/100813enisa_euvd
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-255-01enisa_euvd
- https://www.usa.philips.com/healthcare/about/customer-support/product-securityenisa_euvd
- http://www.securityfocus.com/bid/100813Third Party AdvisoryVDB Entrynvd
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-255-01Third Party AdvisoryUS Government Resourcenvd
Attackers weaponise vulnerabilities like this against internet-exposed services. See what is exposed on your perimeter with a free passive scan.
Free scan