CVE-2014-5406
highpriority · track · 37OT risk · elevatedEPSS 0.012 · 68 %ileEUUSEPSS · scored 2026-10-22
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.
remediations
Vendor fix, mitigation, or workaround guidance, as published by each source.
No vendor fix yet; a mitigation is available.
- MitigationLifeCare PCA Infusion Systemcisa_csaf_ot
Existing PCA Infusion Systems running Version 5.0 can be upgraded to Version 7.0 when it becomes available. Hospira will be retiring older versions of the LifeCare PCA Infusion System, Versions 2 and Versions 3, by the end of the year, 2015.
cvss
Technical severity, 0–10. One row per scoring source.
base score 0–10 · higher = more severe · hover a vector to decode
weaknesses
The CWE weakness classes this flaw belongs to.
affected products (4)
Products and version ranges known to be affected.
- hospira lifecare pcainfusion firmware≤ 5.0nvdUScpe:2.3:o:hospira:lifecare_pcainfusion_firmware:*:*:*:*:*:*:*:*
advisories (1)
Vendor and ecosystem advisories describing this.
- ICSA-15-125-01BHospira LifeCare PCA Infusion System Vulnerabilities2015-02-05icsa · cisa_csaf_otUS
This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov
also known as (3)
Other names and identifiers for this entity.
sources
Which sources assert this record, and when.
references (14)
External citations — patches, advisories, analysis.
- http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htmcnnvd
- http://www.hospira.com/Patchcnnvd
- http://www.securityfocus.com/bid/74476cnnvd
- https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01cnnvd
- https://xs-sniper.com/blog/2015/06/08/hospira-plum-a-infusion-pump-vulnerabilities/cnnvd
- http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htmenisa_euvd
- https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2015/icsa-15-125-01.jsonenisa_euvd
- https://www.cisa.gov/news-events/ics-advisories/icsa-15-125-01enisa_euvd