IDseverity exploitable
CVE-2024-10476US—OT elevated—
Default credentials are used in the above listed BD Diagnostic Solutions products. If exploited, threat actors may be able to access, modify or delete data, including sensitive inf…
CVE-2023-29066USlow3.5—
The FACSChorus software does not properly assign data access privileges for operating system user accounts. A non-administrative OS account can modify information stored in the loc…
CVE-2023-29065USmedium4.3—
The FACSChorus software database can be accessed directly with the privileges of the currently logged-in user. A threat actor with physical access could potentially gain credential…
CVE-2023-29064USmedium4.3OT elevated—
The FACSChorus software contains sensitive information stored in plaintext. A threat actor could gain hardcoded secrets used by the application, which include tokens and passwords…
CVE-2023-29063USlow2.4OT elevated—
The FACSChorus workstation does not prevent physical access to its PCI express (PCIe) slots, which could allow a threat actor to insert a PCI card designed for memory capture. A th…
CVE-2023-29062USlow3.8—
The Operating System hosting the FACSChorus application is configured to allow transmission of hashed user credentials upon user action without adequately validating the identity o…
CVE-2023-29061USmedium5.2OT elevated—
There is no BIOS password on the FACSChorus workstation. A threat actor with physical access to the workstation can potentially exploit this vulnerability to access the BIOS config…
CVE-2023-29060USmedium5.7OT elevated—
The FACSChorus workstation operating system does not restrict what devices can interact with its USB ports. If exploited, a threat actor with physical access to the workstation cou…
CVE-2023-30565USlow3.5OT elevated—
An insecure connection between Systems Manager and CQI Reporter application could expose infusion data to an attacker.
CVE-2023-30564USmedium6.9—
Alaris Systems Manager does not perform input validation during the Device Import Function.
CVE-2023-30563UShigh8.2—
A malicious file could be uploaded into a System Manager User Import Function resulting in a hijacked session.
CVE-2023-30562USmedium6.7OT elevated—
A GRE dataset file within Systems Manager can be tampered with and distributed to PCUs.
CVE-2023-30561USmedium6.1OT elevated—
The data flowing between the PCU and its modules is insecure. A threat actor with physical access could potentially read or modify data by attaching a specially crafted device whil…
CVE-2023-30560USmedium6.8—
The configuration from the PCU can be modified without authentication using physical connection to the PCU.
CVE-2023-30559USmedium5.7OT elevated—
The firmware update package for the wireless card is not properly signed and can be modified.
CVE-2022-43557USmedium5.3OT elevated—
The BD BodyGuard™ infusion pumps specified allow for access through the RS-232 (serial) port interface. If exploited, threat actors with physical access, specialized equipment and…
CVE-2022-40263UShigh7.8OT elevated—
BD Totalys MultiProcessor, versions 1.70 and earlier, contain hardcoded credentials. If exploited, threat actors may be able to access, modify or delete sensitive information, incl…
CVE-2022-30277USmedium5.7—
BD Synapsys™, versions 4.20, 4.20 SR1, and 4.30, contain an insufficient session expiration vulnerability. If exploited, threat actors may be able to access, modify or delete sensi…
CVE-2022-22767UShigh8.8OT elevated—
Specific BD Pyxis™ products were installed with default credentials and may presently still operate with these credentials. There may be scenarios where BD Pyxis™ products are inst…
CVE-2022-22765UShigh7.8OT elevated—
BD Viper LT system, versions 2.0 and later, contains hardcoded credentials. If exploited, threat actors may be able to access, modify or delete sensitive information, including ele…