CVE-2023-25931

mediumPriorität · verfolgen · 27EPSS 0.003 · 19. PerzentilEUUS
grimoire priorityP3 · Track
Exploitable?
—
no known exploitation
Likely / severe?
0.27% (19th) EPSS
CVSS 6.8 · medium
Is it mine?
8 products
Prüfe deine Angriffsfläche
Who says so?
4 sources
cnnvd · enisa_euvd · cisa_csaf_ot · nvd

EPSS · bewertet 2026-10-27

Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.

01

Behebungen

Hersteller-Fix, Mitigation oder Workaround, wie von den Quellen veröffentlicht.

Ein Hersteller-Fix ist verfügbar.

  • BehobenInterStim X Cliniciancisa_csaf_ot

    An app update is available as of February 23, 2023 that will fix the vulnerability.

  • BehobenMicro Cliniciancisa_csaf_ot

    An app update is available as of February 23, 2023 that will fix the vulnerability.

02

cvss

Technischer Schweregrad, 0–10. Eine Zeile je Quelle.

6.8
010

Basiswert 0–10 · höher = schwerwiegender · Vektor überfahren zum Auflösen

v3.16.8mediumCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:Hprimary · NVD / NIST USAusnutzbarkeit 0.9 · Auswirkung 5.9
v3.06.4mediumCVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:Hsecondary · CISA US
v3.16.4mediumCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:Hsecondary · ENISA EU
v3.16.4mediumCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:Hsecondary · security@medtronic.comAusnutzbarkeit 0.9 · Auswirkung 5.5
v3.16.2mediumCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H/E:Usecondary · CVSS-BT (t0sche)
03

SSVC

Eine Handlungsentscheidung — aus Ausnutzung und Auswirkung.

Trackcoordinatorcisa_ssvcUS
04

schwächen

Die CWE-Schwächeklassen, zu denen dies gehört.

05

betroffene Produkte (8)

Betroffene Produkte und Versionsbereiche.

06

Advisories (1)

Hersteller- und Ökosystem-Meldungen dazu.

  • ICSMA-23-061-01Medtronic Micro Clinician and InterStim Apps2023-03-06icsa · cisa_csaf_otUS

    Successful exploitation of this vulnerability could cause the clinician application’s custom password to be reset to default, resulting in unauthorized control of the clinician therapy application.

07

auch bekannt als (3)

Weitere Namen und Kennungen dieser Entität.

08

quellen

Welche Quellen diesen Datensatz belegen und wann.

veröffentlicht 2023-03-01geändert 2026-06-17
10

Referenzen (5)

Externe Belege — Patches, Meldungen, Analysen.

scry

Angreifer nutzen Schwachstellen wie diese gegen im Internet exponierte Dienste aus. Sehen Sie mit einem kostenlosen Passiv-Scan, was exponiert ist.

Kostenloser Scan