CVE-2022-26390

mediumPriorität · verfolgen · 17OT-Risiko · erhöhtEPSS 0.005 · 41. PerzentilEUUS
grimoire priorityP3 · Track
Exploitable?
—
no known exploitation
Likely / severe?
0.47% (41st) EPSS
CVSS 4.2 · medium
Is it mine?
38 products
Prüfe deine Angriffsfläche
Who says so?
3 sources
cnnvd · enisa_euvd · cisa_csaf_ot

EPSS · bewertet 2026-10-27

The Baxter Spectrum Wireless Battery Module (WBM) stores network credentials and PHI (only applicable to Spectrum IQ pumps using auto programming) in unencrypted form. An attacker with physical access to a device that hasn't had all data and settings erased may be able to extract sensitive information.

01

Behebungen

Hersteller-Fix, Mitigation oder Workaround, wie von den Quellen veröffentlicht.

Ein Hersteller-Fix ist verfügbar.

  • BehobenBaxter Spectrum IQcisa_csaf_ot

    According to Baxter, software updates to disable Telnet and FTP (CVE-2022-26392) are in process. Software updates addressing the format string attack (CVE-2022-26393) are included in WBM version 20D30 and all other WBM versions authentication is already available in Spectrum IQ (CVE-2022-26394). Instructions to erase all data and settings on WBMs and pumps before decommissioning and transferring to other facilities (CVE-2022-26390) are in process for incorporation into the Spectrum Operator's Manual.

  • BehobenBaxter Spectrum IQ LVPcisa_csaf_ot

    According to Baxter, software updates to disable Telnet and FTP (CVE-2022-26392) are in process. Software updates addressing the format string attack (CVE-2022-26393) are included in WBM version 20D30 and all other WBM versions authentication is already available in Spectrum IQ (CVE-2022-26394). Instructions to erase all data and settings on WBMs and pumps before decommissioning and transferring to other facilities (CVE-2022-26390) are in process for incorporation into the Spectrum Operator's Manual.

  • BehobenSigma Spectrumcisa_csaf_ot

    According to Baxter, software updates to disable Telnet and FTP (CVE-2022-26392) are in process. Software updates addressing the format string attack (CVE-2022-26393) are included in WBM version 20D30 and all other WBM versions authentication is already available in Spectrum IQ (CVE-2022-26394). Instructions to erase all data and settings on WBMs and pumps before decommissioning and transferring to other facilities (CVE-2022-26390) are in process for incorporation into the Spectrum Operator's Manual.

  • BehobenSigma Spectrum LVPcisa_csaf_ot

    According to Baxter, software updates to disable Telnet and FTP (CVE-2022-26392) are in process. Software updates addressing the format string attack (CVE-2022-26393) are included in WBM version 20D30 and all other WBM versions authentication is already available in Spectrum IQ (CVE-2022-26394). Instructions to erase all data and settings on WBMs and pumps before decommissioning and transferring to other facilities (CVE-2022-26390) are in process for incorporation into the Spectrum Operator's Manual.

  • MitigationBaxter Spectrum IQcisa_csaf_ot

    As a last resort, users may disable wireless operation of the pump; the Spectrum Infusion System was designed to operate without network access. This action would impact an organization’s ability to rapidly deploy drug library (formulary) updates to their pumps.

  • MitigationBaxter Spectrum IQ LVPcisa_csaf_ot

    As a last resort, users may disable wireless operation of the pump; the Spectrum Infusion System was designed to operate without network access. This action would impact an organization’s ability to rapidly deploy drug library (formulary) updates to their pumps.

  • MitigationSigma Spectrumcisa_csaf_ot

    As a last resort, users may disable wireless operation of the pump; the Spectrum Infusion System was designed to operate without network access. This action would impact an organization’s ability to rapidly deploy drug library (formulary) updates to their pumps.

  • MitigationSigma Spectrum LVPcisa_csaf_ot

    As a last resort, users may disable wireless operation of the pump; the Spectrum Infusion System was designed to operate without network access. This action would impact an organization’s ability to rapidly deploy drug library (formulary) updates to their pumps.

02

cvss

Technischer Schweregrad, 0–10. Eine Zeile je Quelle.

4.2
010

Basiswert 0–10 · höher = schwerwiegender · Vektor überfahren zum Auflösen⚠ Quellen uneinig · low–medium · 3.9–4.2

v3.14.2mediumCVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:Nprimary · NVD / NIST USAusnutzbarkeit 0.5 · Auswirkung 3.6
v3.04.2mediumCVSS:3.0/AV:P/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:Nsecondary · CISA US
v3.14.2mediumCVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:Nsecondary · ENISA EU
v3.14.2mediumCVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:Nsecondary · productsecurity@baxter.comAusnutzbarkeit 0.5 · Auswirkung 3.6
v3.13.9lowCVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N/E:Usecondary · CVSS-BT (t0sche)
03

schwächen

Die CWE-Schwächeklassen, zu denen dies gehört.

04

betroffene Produkte (38)

Betroffene Produkte und Versionsbereiche.

20 von 38 angezeigt
05

Advisories (1)

Hersteller- und Ökosystem-Meldungen dazu.

  • ICSMA-22-251-01Baxter Sigma Spectrum Infusion Pump (Update A)2022-09-08icsa · cisa_csaf_otUS

    Successful exploitation of these vulnerabilities could result in access to sensitive data and alteration of system configuration.

06

auch bekannt als (3)

Weitere Namen und Kennungen dieser Entität.

07

quellen

Welche Quellen diesen Datensatz belegen und wann.

veröffentlicht 2022-09-09geändert 2026-06-17
09

Referenzen (8)

Externe Belege — Patches, Meldungen, Analysen.

scry

Angreifer nutzen Schwachstellen wie diese gegen im Internet exponierte Dienste aus. Sehen Sie mit einem kostenlosen Passiv-Scan, was exponiert ist.

Kostenloser Scan