CVE-2021-38394

mediumPriorität · verfolgen · 26EPSS 0.002 · 14. PerzentilEUUS
grimoire priorityP3 · Track
Exploitable?
—
no known exploitation
Likely / severe?
0.23% (14th) EPSS
CVSS 6.9 · medium
Is it mine?
4 products
Prüfe deine Angriffsfläche
Who says so?
4 sources
cnnvd · enisa_euvd · cisa_csaf_ot · nvd

EPSS · bewertet 2026-10-27

An attacker with physical access to the device can extract the binary that checks for the hardware key and reverse engineer it, which could be used to create a physical duplicate of a valid hardware key. The hardware key allows access to special settings when inserted.

01

Behebungen

Hersteller-Fix, Mitigation oder Workaround, wie von den Quellen veröffentlicht.

Ein Hersteller-Fix ist verfügbar.

  • BehobenZOOM LATITUDE Programmer/Recorder/Monitor (PRM) Modelcisa_csaf_ot

    Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security, the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120.

02

cvss

Technischer Schweregrad, 0–10. Eine Zeile je Quelle.

6.9
010

Basiswert 0–10 · höher = schwerwiegender · Vektor überfahren zum Auflösen

v2.06.9mediumAV:L/AC:M/Au:N/C:C/I:C/A:Cprimary · NVD / NIST USAusnutzbarkeit 3.4 · Auswirkung 10
v3.16.4mediumCVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:Hprimary · NVD / NIST USAusnutzbarkeit 0.5 · Auswirkung 5.9
v3.06.2mediumCVSS:3.0/AV:P/AC:H/PR:N/UI:R/S:C/C:L/I:H/A:Lsecondary · CISA US
v3.16.2mediumCVSS:3.1/AV:P/AC:H/PR:N/UI:R/S:C/C:L/I:H/A:Lsecondary · ENISA EU
v3.16.2mediumCVSS:3.1/AV:P/AC:H/PR:N/UI:R/S:C/C:L/I:H/A:Lsecondary · ics-cert@hq.dhs.gov USAusnutzbarkeit 0.4 · Auswirkung 5.3
v3.15.9mediumCVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H/E:Usecondary · CVSS-BT (t0sche)
03

schwächen

Die CWE-Schwächeklassen, zu denen dies gehört.

04

betroffene Produkte (4)

Betroffene Produkte und Versionsbereiche.

05

Advisories (1)

Hersteller- und Ökosystem-Meldungen dazu.

  • ICSMA-21-273-01Boston Scientific Zoom Latitude2021-09-30icsa · cisa_csaf_otUS

    Successful exploitation of these vulnerabilities may allow an attacker with physical access to the affected device to obtain patient protected health information (PHI), and/or compromise the integrity of the device. The affected device is not network connected and does not contain hardware to be network connected.

06

auch bekannt als (3)

Weitere Namen und Kennungen dieser Entität.

07

quellen

Welche Quellen diesen Datensatz belegen und wann.

veröffentlicht 2021-10-04geändert 2026-06-17
08

Referenzen (6)

Externe Belege — Patches, Meldungen, Analysen.

scry

Angreifer nutzen Schwachstellen wie diese gegen im Internet exponierte Dienste aus. Sehen Sie mit einem kostenlosen Passiv-Scan, was exponiert ist.

Kostenloser Scan