CVE-2019-13535

lowPriorität · verfolgen · 18EPSS 0.004 · 30. PerzentilEUUS
grimoire priorityP3 · Track
Exploitable?
—
no known exploitation
Likely / severe?
0.37% (30th) EPSS
CVSS 2.1 · low
Is it mine?
12 products
Prüfe deine Angriffsfläche
Who says so?
4 sources
cnnvd · enisa_euvd · cisa_csaf_ot · nvd

EPSS · bewertet 2026-10-27

In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States) version 1.20.2 and lower, the RFID security mechanism does not apply read protection, allowing for full read access of the RFID security mechanism data.

01

Behebungen

Hersteller-Fix, Mitigation oder Workaround, wie von den Quellen veröffentlicht.

Ein Hersteller-Fix ist verfügbar.

  • BehobenValleylab FT10 Energy Platform (VLFT10GEN)cisa_csaf_ot

    Medtronic has released additional patient focused information at the following location:

  • BehobenValleylab LS10 Energy Platform (VLLS10GEN)cisa_csaf_ot

    Medtronic has released additional patient focused information at the following location:

02

cvss

Technischer Schweregrad, 0–10. Eine Zeile je Quelle.

4.6
010

Basiswert 0–10 · höher = schwerwiegender · Vektor überfahren zum Auflösen⚠ Quellen uneinig · low–medium · 2.1–4.6

v2.02.1lowAV:L/AC:L/Au:N/C:P/I:N/A:Nprimary · NVD / NIST USAusnutzbarkeit 3.9 · Auswirkung 2.9
v3.14.6mediumCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:Nprimary · NVD / NIST USAusnutzbarkeit 0.9 · Auswirkung 3.6
v3.04.6mediumCVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:Nsecondary · CISA US
v3.14.6mediumCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:Nsecondary · ENISA EU
v3.14.6mediumCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:Nsecondary · ics-cert@hq.dhs.gov USAusnutzbarkeit 0.9 · Auswirkung 3.6
v3.14.2mediumCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N/E:Usecondary · CVSS-BT (t0sche)
03

schwächen

Die CWE-Schwächeklassen, zu denen dies gehört.

04

betroffene Produkte (12)

Betroffene Produkte und Versionsbereiche.

05

Advisories (1)

Hersteller- und Ökosystem-Meldungen dazu.

  • ICSMA-19-311-01Medtronic Valleylab FT10 and LS102019-11-07icsa · cisa_csaf_otUS

    Successful exploitation of these vulnerabilities may allow an attacker to connect inauthentic instruments to the affected products by spoofing RFID security mechanisms. This may lead to a loss of performance integrity and platform availability due to incorrect identification of instrument and associated parameters.

06

auch bekannt als (3)

Weitere Namen und Kennungen dieser Entität.

07

quellen

Welche Quellen diesen Datensatz belegen und wann.

veröffentlicht 2019-11-08geändert 2026-06-17
09

Referenzen (10)

Externe Belege — Patches, Meldungen, Analysen.

scry

Angreifer nutzen Schwachstellen wie diese gegen im Internet exponierte Dienste aus. Sehen Sie mit einem kostenlosen Passiv-Scan, was exponiert ist.

Kostenloser Scan