CVE-2018-10596

mediumPriorität · verfolgen · 32EPSS 0.013 · 71. PerzentilEUUS
grimoire priorityP3 · Track
Exploitable?
—
no known exploitation
Likely / severe?
1.32% (71st) EPSS
CVSS 5.2 · medium
Is it mine?
7 products
Prüfe deine Angriffsfläche
Who says so?
4 sources
cnnvd · enisa_euvd · cisa_csaf_ot · nvd

EPSS · bewertet 2026-10-26

Medtronic 2090 CareLink Programmer uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications.

01

Behebungen

Hersteller-Fix, Mitigation oder Workaround, wie von den Quellen veröffentlicht.

Ein Hersteller-Fix ist verfügbar.

  • Behoben2090 CareLink Programmercisa_csaf_ot

    After additional review and risk evaluation of the affected products, Medtronic has disabled the network-based software update mechanism, including both the VPN and the HTTP subservices, as an immediate security mitigation. Users should not attempt to update the affected products over the network as this update mechanism is vulnerable to the attack described in section 4.2.3. Medtronic will continue to implement and deploy increased security protections and mitigations to address the vulnerabilities in this advisory.

  • Behoben29901 Encore Programmercisa_csaf_ot

    After additional review and risk evaluation of the affected products, Medtronic has disabled the network-based software update mechanism, including both the VPN and the HTTP subservices, as an immediate security mitigation. Users should not attempt to update the affected products over the network as this update mechanism is vulnerable to the attack described in section 4.2.3. Medtronic will continue to implement and deploy increased security protections and mitigations to address the vulnerabilities in this advisory.

02

cvss

Technischer Schweregrad, 0–10. Eine Zeile je Quelle.

8.0
010

Basiswert 0–10 · höher = schwerwiegender · Vektor überfahren zum Auflösen⚠ Quellen uneinig · medium–high · 5.2–8

v2.05.2mediumAV:A/AC:L/Au:S/C:P/I:P/A:Pprimary · NVD / NIST USAusnutzbarkeit 5.1 · Auswirkung 6.4
v3.08highCVSS:3.0/AV:A/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:Hprimary · NVD / NIST USAusnutzbarkeit 2.1 · Auswirkung 5.9
v3.07.1highCVSS:3.0/AV:N/AC:H/PR:L/UI:R/S:U/C:H/I:H/A:Hsecondary · CISA US
v3.17.1highCVSS:3.1/AV:N/AC:H/PR:L/UI:R/S:U/C:H/I:H/A:Hsecondary · ENISA EU
v3.17.1highCVSS:3.1/AV:N/AC:H/PR:L/UI:R/S:U/C:H/I:H/A:Hsecondary · ics-cert@hq.dhs.gov USAusnutzbarkeit 1.2 · Auswirkung 5.9
v3.16.5mediumCVSS:3.1/AV:N/AC:H/PR:L/UI:R/S:U/C:H/I:H/A:H/E:Usecondary · CVSS-BT (t0sche)
03

schwächen

Die CWE-Schwächeklassen, zu denen dies gehört.

04

betroffene Produkte (7)

Betroffene Produkte und Versionsbereiche.

05

Advisories (1)

Hersteller- und Ökosystem-Meldungen dazu.

  • ICSMA-18-058-01Medtronic 2090 Carelink Programmer Vulnerabilities (Update C)2018-02-27icsa · cisa_csaf_otUS

    Successful exploitation of these vulnerabilities may allow an attacker with physical access to a 2090 Programmer to obtain per-product credentials to the software deployment network. These credentials grant access to the software deployment network, but access is limited to read-only versions of device software applications.

06

auch bekannt als (3)

Weitere Namen und Kennungen dieser Entität.

07

quellen

Welche Quellen diesen Datensatz belegen und wann.

veröffentlicht 2018-07-03geändert 2026-06-17
09

Referenzen (10)

Externe Belege — Patches, Meldungen, Analysen.

scry

Angreifer nutzen Schwachstellen wie diese gegen im Internet exponierte Dienste aus. Sehen Sie mit einem kostenlosen Passiv-Scan, was exponiert ist.

Kostenloser Scan