CVE-2017-5149

mediumPriorität · verfolgen · 36EPSS 0.012 · 69. PerzentilEUUS
grimoire priorityP3 · Track
Exploitable?
—
no known exploitation
Likely / severe?
1.24% (69th) EPSS
CVSS 6.8 · medium
Is it mine?
8 products
Prüfe deine Angriffsfläche
Who says so?
4 sources
cnnvd · enisa_euvd · cisa_csaf_ot · nvd

EPSS · bewertet 2026-10-27

An issue was discovered in St. Jude Medical Merlin@home, versions prior to Version 8.2.2 (RF models: EX1150; Inductive models: EX1100; and Inductive models: EX1100 with MerlinOnDemand capability). The identities of the endpoints for the communication channel between the transmitter and St. Jude Medical's web site, Merlin.net, are not verified. This may allow a man-in-the-middle attacker to access or influence communications between the identified endpoints.

01

Behebungen

Hersteller-Fix, Mitigation oder Workaround, wie von den Quellen veröffentlicht.

Noch kein Hersteller-Fix; eine Mitigation ist verfügbar.

  • MitigationInductive models EX1100cisa_csaf_ot

    The new version of the transmitter software, Version 8.2.2, will be automatically updated over a period of several months, when all models of the Merlin@home transmitters are connected to an Ethernet, WiFi, cellular network, or a landline. St. Jude Medical recommends that users keep Merlin@home transmitters powered and connected at all times to receive this update and future updates.

  • MitigationInductive models EX1100 with MerlinOnDemand capabilitycisa_csaf_ot

    The new version of the transmitter software, Version 8.2.2, will be automatically updated over a period of several months, when all models of the Merlin@home transmitters are connected to an Ethernet, WiFi, cellular network, or a landline. St. Jude Medical recommends that users keep Merlin@home transmitters powered and connected at all times to receive this update and future updates.

  • MitigationRF models EX1150cisa_csaf_ot

    The new version of the transmitter software, Version 8.2.2, will be automatically updated over a period of several months, when all models of the Merlin@home transmitters are connected to an Ethernet, WiFi, cellular network, or a landline. St. Jude Medical recommends that users keep Merlin@home transmitters powered and connected at all times to receive this update and future updates.

02

cvss

Technischer Schweregrad, 0–10. Eine Zeile je Quelle.

8.9
010

Basiswert 0–10 · höher = schwerwiegender · Vektor überfahren zum Auflösen⚠ Quellen uneinig · medium–high · 6.8–8.9

v2.06.8mediumAV:N/AC:M/Au:N/C:P/I:P/A:Pprimary · NVD / NIST USAusnutzbarkeit 8.6 · Auswirkung 6.4
v3.18.9highCVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:L/I:H/A:Hprimary · NVD / NIST USAusnutzbarkeit 2.2 · Auswirkung 6
v3.08.9highCVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:C/C:L/I:H/A:Hsecondary · CISA US
v3.18.1highCVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:L/I:H/A:H/E:Usecondary · CVSS-BT (t0sche)
03

schwächen

Die CWE-Schwächeklassen, zu denen dies gehört.

04

betroffene Produkte (8)

Betroffene Produkte und Versionsbereiche.

05

Advisories (1)

Hersteller- und Ökosystem-Meldungen dazu.

  • ICSMA-17-009-01AICSMA-17-009-01A_St. Jude Merlin@home Transmitter Vulnerability (Update A)2017-01-09icsa · cisa_csaf_otUS

    This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov

06

auch bekannt als (3)

Weitere Namen und Kennungen dieser Entität.

07

quellen

Welche Quellen diesen Datensatz belegen und wann.

veröffentlicht 2017-02-13geändert 2026-06-17
09

Referenzen (7)

Externe Belege — Patches, Meldungen, Analysen.

scry

Angreifer nutzen Schwachstellen wie diese gegen im Internet exponierte Dienste aus. Sehen Sie mit einem kostenlosen Passiv-Scan, was exponiert ist.

Kostenloser Scan