CVE-2017-12716
lowPriorität · verfolgen · 26OT-Risiko · erhöhtEPSS 0.002 · 15. PerzentilEUUSEPSS · bewertet 2026-10-27
Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
Behebungen
Hersteller-Fix, Mitigation oder Workaround, wie von den Quellen veröffentlicht.
Noch kein Hersteller-Fix; eine Mitigation ist verfügbar.
- MitigationAccent MRIcisa_csaf_ot
The pacemaker firmware update will implement RF wake-up protections and limit the commands that can be issued to pacemakers via RF communications. Additionally the updated pacemaker firmware will prevent unencrypted transmission of patient information (Accent and Anthem only). The firmware update can be applied to an implanted pacemaker via the Merlin PCS Programmer by a healthcare provider. It is recommended that healthcare providers discuss this update with their patients and carefully consider the potential risk of a cybersecurity attack along with the risk of performing a firmware update. Implementation of the firmware update is to be determined based on the physician's professional judgment and patient management considerations. Pacemakers manufactured beginning August 28, 2017, will have this update preloaded on devices.
- MitigationAccent/Anthemcisa_csaf_ot
The pacemaker firmware update will implement RF wake-up protections and limit the commands that can be issued to pacemakers via RF communications. Additionally the updated pacemaker firmware will prevent unencrypted transmission of patient information (Accent and Anthem only). The firmware update can be applied to an implanted pacemaker via the Merlin PCS Programmer by a healthcare provider. It is recommended that healthcare providers discuss this update with their patients and carefully consider the potential risk of a cybersecurity attack along with the risk of performing a firmware update. Implementation of the firmware update is to be determined based on the physician's professional judgment and patient management considerations. Pacemakers manufactured beginning August 28, 2017, will have this update preloaded on devices.
- MitigationAssurity MRIcisa_csaf_ot
The pacemaker firmware update will implement RF wake-up protections and limit the commands that can be issued to pacemakers via RF communications. Additionally the updated pacemaker firmware will prevent unencrypted transmission of patient information (Accent and Anthem only). The firmware update can be applied to an implanted pacemaker via the Merlin PCS Programmer by a healthcare provider. It is recommended that healthcare providers discuss this update with their patients and carefully consider the potential risk of a cybersecurity attack along with the risk of performing a firmware update. Implementation of the firmware update is to be determined based on the physician's professional judgment and patient management considerations. Pacemakers manufactured beginning August 28, 2017, will have this update preloaded on devices.
- MitigationAssurity/Allurecisa_csaf_ot
The pacemaker firmware update will implement RF wake-up protections and limit the commands that can be issued to pacemakers via RF communications. Additionally the updated pacemaker firmware will prevent unencrypted transmission of patient information (Accent and Anthem only). The firmware update can be applied to an implanted pacemaker via the Merlin PCS Programmer by a healthcare provider. It is recommended that healthcare providers discuss this update with their patients and carefully consider the potential risk of a cybersecurity attack along with the risk of performing a firmware update. Implementation of the firmware update is to be determined based on the physician's professional judgment and patient management considerations. Pacemakers manufactured beginning August 28, 2017, will have this update preloaded on devices.
cvss
Technischer Schweregrad, 0–10. Eine Zeile je Quelle.
Basiswert 0–10 · höher = schwerwiegender · Vektor überfahren zum Auflösen⚠ Quellen uneinig · low–medium · 3.1–6.5
schwächen
Die CWE-Schwächeklassen, zu denen dies gehört.
betroffene Produkte (13)
Betroffene Produkte und Versionsbereiche.
- Abbott Laboratories Accent and Anthem≤ All versions of pacemakers manufactured prior to August 28, 2017enisa_euvdEU
Advisories (1)
Hersteller- und Ökosystem-Meldungen dazu.
- ICSMA-17-241-01ICSMA-17-241-01_Abbott Laboratories ' Accent/Anthem, Accent MRI, Assurity/Allure, and Assurity MRI Pacemaker Vulnerabilities2017-08-29icsa · cisa_csaf_otUS
This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov
auch bekannt als (3)
Weitere Namen und Kennungen dieser Entität.
quellen
Welche Quellen diesen Datensatz belegen und wann.
Referenzen (7)
Externe Belege — Patches, Meldungen, Analysen.
- http://abbott.mediaroom.com/2017-08-29-Abbott-issues-new-updates-for-implanted-cardiac-devicesPatchcnnvd
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01cnnvd
- https://www.securityfocus.com/bid/100523cnnvd
- http://www.securityfocus.com/bid/100523enisa_euvd
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01enisa_euvd
- http://www.securityfocus.com/bid/100523Third Party AdvisoryVDB Entrynvd
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01Third Party AdvisoryUS Government Resourcenvd
Angreifer nutzen Schwachstellen wie diese gegen im Internet exponierte Dienste aus. Sehen Sie mit einem kostenlosen Passiv-Scan, was exponiert ist.
Kostenloser Scan