ICSA-15-337-02

icsaUScisa_csaf_ot

Hospira Multiple Products Buffer Overflow Vulnerability

Advisory ansehen veröffentlicht 2015-09-05

This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov

01

beschreibt (1)

Zugehörige CVEs.

IDSchweregrad ausnutzbar
CVE-2015-7909UShigh7.3—
Stack-based buffer overflow in Hospira Communication Engine (CE) before 1.2 in LifeCare PCA Infusion System 5.07, Plum A+ Infusion System 13.40, and Plum A+3 Infusion System 13.40…
02

Referenzen

Externe Belege — Patches, Meldungen, Analysen.

03

Behebungen (3)

  • MitigationLifeCare PCA Infusion System (running CE version 1.0 or earlier)

    Update to CE version 1.2 or later.

  • MitigationPlum A+ Infusion System (running CE version 1.0 or earlier)

    Update to CE version 1.2 or later.

  • MitigationPlum A+3 Infusion System (running CE version 1.0 or earlier)

    Update to CE version 1.2 or later.