Hospira Multiple Products Buffer Overflow Vulnerability
Advisory ansehen veröffentlicht 2015-09-05
This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov
01
beschreibt (1)
Zugehörige CVEs.
| ID | Schweregrad | ausnutzbar | ||
|---|---|---|---|---|
| CVE-2015-7909US | high7.3 | — | ||
Stack-based buffer overflow in Hospira Communication Engine (CE) before 1.2 in LifeCare PCA Infusion System 5.07, Plum A+ Infusion System 13.40, and Plum A+3 Infusion System 13.40… | ||||
02
Referenzen
Externe Belege — Patches, Meldungen, Analysen.
- ICS Advisory ICSA-15-337-02 JSONself
- Recommended Practicesexternal
- ICS Advisory ICSA-15-337-02 - Web Versionself
- Recommended Practicesexternal
- Recommended Practicesexternal
- Recommended Practicesexternal
- Recommended Practicesexternal
- Recommended Practicesexternal
03
Behebungen (3)
- MitigationLifeCare PCA Infusion System (running CE version 1.0 or earlier)
Update to CE version 1.2 or later.
- MitigationPlum A+ Infusion System (running CE version 1.0 or earlier)
Update to CE version 1.2 or later.
- MitigationPlum A+3 Infusion System (running CE version 1.0 or earlier)
Update to CE version 1.2 or later.