Animas OneTouch Ping Insulin Pump Vulnerabilities
This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov
beschreibt (3)
Zugehörige CVEs.
| ID | Schweregrad | ausnutzbar | ||
|---|---|---|---|---|
| CVE-2016-5086US | critical9.8OT erhöht | — | ||
Johnson & Johnson Animas OneTouch Ping devices allow remote attackers to bypass authentication via replay attacks. | ||||
| CVE-2016-5085US | high7.5OT erhöht | — | ||
Johnson & Johnson Animas OneTouch Ping devices do not properly generate random numbers, which makes it easier for remote attackers to spoof meters by sniffing the network and then… | ||||
| CVE-2016-5084US | high7.5OT erhöht | — | ||
Johnson & Johnson Animas OneTouch Ping devices do not use encryption for certain data, which might allow remote attackers to obtain sensitive information by sniffing the network. | ||||
Referenzen
Externe Belege — Patches, Meldungen, Analysen.
- ICS Advisory ICSMA-16-279-01 JSONself
- Recommended Practicesexternal
- ICS Advisory ICSMA-16-279-01 - Web Versionself
- Recommended Practicesexternal
- Recommended Practicesexternal
- Recommended Practicesexternal
- Recommended Practicesexternal
- Recommended Practicesexternal
Behebungen (3)
- MitigationOneTouch Ping insulin pump system
Animas also suggests turning on the Vibrating Alert feature of the OneTouch Ping system, as described in Chapter 4 of Section I. This notifies the user that a bolus dose is being initiated by the meter remote, which gives the patient the option of canceling the bolus.
- MitigationOneTouch Ping insulin pump system
Animas also suggests turning on the Vibrating Alert feature of the OneTouch Ping system, as described in Chapter 4 of Section I. This notifies the user that a bolus dose is being initiated by the meter remote, which gives the patient the option of canceling the bolus.
- MitigationOneTouch Ping insulin pump system
Animas also suggests turning on the Vibrating Alert feature of the OneTouch Ping system, as described in Chapter 4 of Section I. This notifies the user that a bolus dose is being initiated by the meter remote, which gives the patient the option of canceling the bolus.