ICSMA-24-352-01

icsaUScisa_csaf_ot

BD Diagnostic Solutions Products (Update A)

Advisory ansehen veröffentlicht 2024-12-17

Successful exploitation of this vulnerability could allow an attacker to use default credentials to access, modify, or delete sensitive data, which could impact the availability of the system or cause a system shutdown.

01

beschreibt (1)

Zugehörige CVEs.

IDSchweregrad ausnutzbar
CVE-2024-10476US—OT erhöht—
Default credentials are used in the above listed BD Diagnostic Solutions products. If exploited, threat actors may be able to access, modify or delete data, including sensitive inf…
02

Referenzen

Externe Belege — Patches, Meldungen, Analysen.

03

Behebungen (7)

  • MitigationBD BACTEC Blood Culture System

    BD has already communicated to users with affected products and is working with them to update default credentials on affected products. For this vulnerability to be exploited, a threat actor will need direct access, whether logical or physical, into the clinical setting.

  • MitigationBD COR System

    BD has already communicated to users with affected products and is working with them to update default credentials on affected products. For this vulnerability to be exploited, a threat actor will need direct access, whether logical or physical, into the clinical setting.

  • MitigationBD EpiCenter Microbiology Data Management System

    BD has already communicated to users with affected products and is working with them to update default credentials on affected products. For this vulnerability to be exploited, a threat actor will need direct access, whether logical or physical, into the clinical setting.

  • MitigationBD MAX System

    BD has already communicated to users with affected products and is working with them to update default credentials on affected products. For this vulnerability to be exploited, a threat actor will need direct access, whether logical or physical, into the clinical setting.

  • MitigationBD Phoenix M50 Automated Microbiology System

    BD has already communicated to users with affected products and is working with them to update default credentials on affected products. For this vulnerability to be exploited, a threat actor will need direct access, whether logical or physical, into the clinical setting.

  • MitigationBD Synapsys Informatics Solution

    BD has already communicated to users with affected products and is working with them to update default credentials on affected products. For this vulnerability to be exploited, a threat actor will need direct access, whether logical or physical, into the clinical setting.

  • BehobenBD Synapsys Informatics Solution

    Note: BD Synapsys Informatics Solution is only in scope of this vulnerability when installed on a NUC server. BD Synapsys Informatics Solution installed on a user-provided virtual machine or on the BD Kiestra SCU hardware is not in scope.