Baxter Life2000 Ventilation System
Successful exploitation of these vulnerabilities could lead to information disclosure and/or disruption of the device's function without detection.
beschreibt (9)
Zugehörige CVEs.
| ID | Schweregrad | ausnutzbar | ||
|---|---|---|---|---|
| CVE-2024-48974US | —OT erhöht | — | ||
The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's co… | ||||
| CVE-2024-48973US | — | — | ||
The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see… | ||||
| CVE-2024-48971US | —OT erhöht | — | ||
The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventil… | ||||
| CVE-2024-48970US | — | — | ||
The ventilator's microcontroller lacks memory protection. An attacker could connect to the internal JTAG interface and read or write to flash memory using an off-the-shelf debuggin… | ||||
| CVE-2024-48967US | — | — | ||
The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with acce… | ||||
| CVE-2024-48966US | —OT erhöht | — | ||
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are in… | ||||
| CVE-2024-9834US | —OT erhöht | — | ||
Improper data protection on the ventilator's serial interface could allow an attacker to send and receive messages that result in unauthorized disclosure of information and/or have… | ||||
| CVE-2024-9832US | — | — | ||
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force atta… | ||||
| CVE-2020-8004US | high7.5 | Exploit | ||
STMicroelectronics STM32F1 devices have Incorrect Access Control. | ||||
Referenzen
Externe Belege — Patches, Meldungen, Analysen.
- ICS Advisory ICSMA-24-319-01 JSONself
- Recommended Practicesexternal
- ICSA Advisory ICSMA-24-319-01 - Web Versionself
- Recommended Practicesexternal
- Recommended Practicesexternal
- Recommended Practicesexternal
- Recommended Practicesexternal
- Recommended Practicesexternal
Behebungen (18)
- MitigationLife2000 Ventilation System
Baxter is unaware of any exploitation of these vulnerabilities and/or the compromise of personal or health data.
- MitigationLife2000 Ventilation System
Baxter is unaware of any exploitation of these vulnerabilities and/or the compromise of personal or health data.
- MitigationLife2000 Ventilation System
Baxter is unaware of any exploitation of these vulnerabilities and/or the compromise of personal or health data.
- MitigationLife2000 Ventilation System
Baxter is unaware of any exploitation of these vulnerabilities and/or the compromise of personal or health data.
- MitigationLife2000 Ventilation System
Baxter is unaware of any exploitation of these vulnerabilities and/or the compromise of personal or health data.
- MitigationLife2000 Ventilation System
Baxter is unaware of any exploitation of these vulnerabilities and/or the compromise of personal or health data.
- MitigationLife2000 Ventilation System
Baxter is unaware of any exploitation of these vulnerabilities and/or the compromise of personal or health data.
- MitigationLife2000 Ventilation System
Baxter is unaware of any exploitation of these vulnerabilities and/or the compromise of personal or health data.
- MitigationLife2000 Ventilation System
Baxter is unaware of any exploitation of these vulnerabilities and/or the compromise of personal or health data.
- BehobenLife2000 Ventilation System
Baxter recommends that users of the Life2000 Ventilation System not leave their ventilators unattended in public or unsecured areas. Maintaining physical possession and control of the ventilator reduces the likelihood of a malicious actor gaining access to the device.
- BehobenLife2000 Ventilation System
Baxter recommends that users of the Life2000 Ventilation System not leave their ventilators unattended in public or unsecured areas. Maintaining physical possession and control of the ventilator reduces the likelihood of a malicious actor gaining access to the device.
- BehobenLife2000 Ventilation System
Baxter recommends that users of the Life2000 Ventilation System not leave their ventilators unattended in public or unsecured areas. Maintaining physical possession and control of the ventilator reduces the likelihood of a malicious actor gaining access to the device.
- BehobenLife2000 Ventilation System
Baxter recommends that users of the Life2000 Ventilation System not leave their ventilators unattended in public or unsecured areas. Maintaining physical possession and control of the ventilator reduces the likelihood of a malicious actor gaining access to the device.
- BehobenLife2000 Ventilation System
Baxter recommends that users of the Life2000 Ventilation System not leave their ventilators unattended in public or unsecured areas. Maintaining physical possession and control of the ventilator reduces the likelihood of a malicious actor gaining access to the device.
- BehobenLife2000 Ventilation System
Baxter recommends that users of the Life2000 Ventilation System not leave their ventilators unattended in public or unsecured areas. Maintaining physical possession and control of the ventilator reduces the likelihood of a malicious actor gaining access to the device.
- BehobenLife2000 Ventilation System
Baxter recommends that users of the Life2000 Ventilation System not leave their ventilators unattended in public or unsecured areas. Maintaining physical possession and control of the ventilator reduces the likelihood of a malicious actor gaining access to the device.
- BehobenLife2000 Ventilation System
Baxter recommends that users of the Life2000 Ventilation System not leave their ventilators unattended in public or unsecured areas. Maintaining physical possession and control of the ventilator reduces the likelihood of a malicious actor gaining access to the device.
- BehobenLife2000 Ventilation System
Baxter recommends that users of the Life2000 Ventilation System not leave their ventilators unattended in public or unsecured areas. Maintaining physical possession and control of the ventilator reduces the likelihood of a malicious actor gaining access to the device.