Medtronic Paceart Optima System
Advisory ansehen veröffentlicht 2023-06-29
Successful exploitation of this vulnerability could result in remote code execution or a denial-of-service condition impacting a healthcare delivery organization's Paceart Optima system.
01
beschreibt (1)
Zugehörige CVEs.
| ID | Schweregrad | ausnutzbar | ||
|---|---|---|---|---|
| CVE-2023-31222US | high8.8 | — | ||
Deserialization of untrusted data in Microsoft Messaging Queuing Service in Medtronic's Paceart Optima versions 1.11 and earlier on Windows allows an unauthorized user to impact a… | ||||
02
Referenzen
Externe Belege — Patches, Meldungen, Analysen.
- ICS Advisory ICSMA-23-180-01 JSONself
- Recommended Practicesexternal
- ICSA Advisory ICSMA-23-180-01 - Web Versionself
- Recommended Practicesexternal
- Recommended Practicesexternal
- Recommended Practicesexternal
- Recommended Practicesexternal
- Recommended Practicesexternal
03
Behebungen (2)
- MitigationPaceart Optima
As long as the Paceart Messaging Service remains disabled, the vulnerability will remain mitigated.
- BehobenPaceart Optima
Medtronic recommends updating the Paceart Optima system to v1.12. Contact Medtronic to schedule the update. (mailto:dl.paceartupdate@medtronic.com)
scry
Advisories wie diese benennen die exponierten Produkte, die Angreifer auf Ihrer Angriffsfläche ins Visier nehmen. Sehen Sie mit einem kostenlosen Passiv-Scan, was exponiert ist.