VDE-2020-021

csafDEendress_hauser_psirt

Endress+Hauser: Ecograph T utilizing Webserver firmware version 1.x suffers from improper privilege management

view advisory published 2020-11-19

The affected device has a web-based user interface with a role-based access system. Users with different roles have different write and read privileges. The access system is based on dynamic "tokens".

01

describes (1)

Related CVEs.

IDseverity exploitable
CVE-2020-12495UShigh8.8—
Endress+Hauser Ecograph T (Neutral/Private Label) (RSG35, ORSG35) with Firmware version prior to V2.0.0 is prone to improper privilege management. The affected device has a web-bas…
02

references

External citations — patches, advisories, analysis.

03

remediations (1)

  • Mitigation

    Customers should configure a perimeter firewall to block traffic from untrusted networks and users to the device. These recommendations will be incorporated into the device documentation (operating instructions). Change default password for operator, service and admin account. Endress+Hauser will not change this behavior. Customers are recommended to take the measures for Temporary Fix / Mitigation as described above.