ICSMA-18-219-01

icsaUScisa_csaf_ot

Medtronic MyCareLink 24950 Patient Monitor (Update A)

view advisory published 2018-08-07

Successful exploitation of these vulnerabilities may allow an attacker with physical access to obtain per-product credentials that are utilized to authenticate data uploads and encrypt data at rest. Additionally, an attacker with access to a set of these credentials and additional identifiers can upload invalid data to the Medtronic CareLink network.

01

describes (2)

Related CVEs.

IDseverity exploitable
CVE-2018-10626USmedium4.4OT elevated—
Medtronic MyCareLink Patient Monitor’s update service does not sufficiently verify the authenticity of the data uploaded. An attacker who obtains per-product credentials from the…
CVE-2018-10622UShigh7.1—
Medtronic MyCareLink Patient Monitor uses per-product credentials that are stored in a recoverable format. An attacker can use these credentials for network authentication.
02

references

External citations — patches, advisories, analysis.

03

remediations (4)

  • Mitigation24950 MyCareLink Monitor

    Maintain good physical control over the home monitor.

  • Mitigation24950 MyCareLink Monitor

    Maintain good physical control over the home monitor.

  • Mitigation24952 MyCareLink Monitor

    Maintain good physical control over the home monitor.

  • Mitigation24952 MyCareLink Monitor

    Maintain good physical control over the home monitor.