ICSMA-25-205-01

icsaUScisa_csaf_ot

Medtronic MyCareLink Patient Monitor (Update A)

view advisory published 2025-07-24

Successful exploitation of these vulnerabilities could lead to system compromise, unauthorized access to sensitive data, and manipulation of the monitor's functionality.

01

describes (6)

Related CVEs.

IDseverity exploitable
CVE-2025-4397US——
Medtronic MyCareLink Patient Monitor uses per-product credentials that are stored in a recoverable format. An attacker can use these credentials to modify encrypted drive data.
CVE-2025-4386US——
Medtronic MyCareLink Patient Monitor has an internal serial interface, which allows an attacker with physical access to access a login prompt via a UART terminal.​
CVE-2025-4395US——
Medtronic MyCareLink Patient Monitor has a built-in user account with an empty password, which allows an attacker with physical access to log in with no password and access modify…
CVE-2025-4394US—OT elevated—
Medtronic MyCareLink Patient Monitor uses an unencrypted filesystem on internal storage, which allows an attacker with physical access to read and modify files. This issue affect…
CVE-2025-4393US——
Medtronic MyCareLink Patient Monitor has an internal service that deserializes data, which allows a local attacker to interact with the service by crafting a binary payload to cras…
CVE-2018-10622UShigh7.1—
Medtronic MyCareLink Patient Monitor uses per-product credentials that are stored in a recoverable format. An attacker can use these credentials for network authentication.
02

references

External citations — patches, advisories, analysis.

03

remediations (12)

  • MitigationMyCareLink Patient Monitor model 24950

    Medtronic recommends the following actions:

  • MitigationMyCareLink Patient Monitor model 24950

    Medtronic recommends the following actions:

  • MitigationMyCareLink Patient Monitor model 24950

    Medtronic recommends the following actions:

  • MitigationMyCareLink Patient Monitor model 24950

    Medtronic recommends the following actions:

  • MitigationMyCareLink Patient Monitor model 24950

    Medtronic recommends the following actions:

  • MitigationMyCareLink Patient Monitor model 24950

    Medtronic recommends the following actions:

  • MitigationMyCareLink Patient Monitor model 24952

    Medtronic recommends the following actions:

  • MitigationMyCareLink Patient Monitor model 24952

    Medtronic recommends the following actions:

  • MitigationMyCareLink Patient Monitor model 24952

    Medtronic recommends the following actions:

  • MitigationMyCareLink Patient Monitor model 24952

    Medtronic recommends the following actions:

  • MitigationMyCareLink Patient Monitor model 24952

    Medtronic recommends the following actions:

  • MitigationMyCareLink Patient Monitor model 24952

    Medtronic recommends the following actions: