Hospira LifeCare PCA Infusion System Vulnerabilities
This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov
describes (1)
Related CVEs.
| ID | severity | exploitable | ||
|---|---|---|---|---|
| CVE-2015-3459US | high10.0 | — | ||
The communication module on the Hospira LifeCare PCA Infusion System before 7.0 does not require authentication for root TELNET sessions, which allows remote attackers to modify th… | ||||
references
External citations — patches, advisories, analysis.
- ICS Advisory ICSA-15-125-01 JSONself
- Recommended Practicesexternal
- ICS Advisory ICSA-15-125-01 - Web Versionself
- Recommended Practicesexternal
- Recommended Practicesexternal
- Recommended Practicesexternal
- Recommended Practicesexternal
- Recommended Practicesexternal
remediations (1)
- MitigationLifeCare PCA Infusion System
CS-CERT has been working with Hospira since May 2014 to address the vulnerabilities in the LifeCare PCA Infusion System. Hospira has developed a new version of the PCS Infusion System, Version 7.0 that addresses the identified vulnerabilities. According to Hospira, Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access. Existing PCA Infusion Systems running Version 5.0 can be upgraded to Version 7.0 when it becomes available. Hospira’s Version 7.0 is being reviewed by the FDA prior to its release. The release date for Version 7.0 of the LifeCare PCA Infusion System has not been determined. For additional information about Hospira’s new release, contact Hospira’s technical support at 1‑800-241-4002. ICS-CERT encourages asset owners to take defensive measures to protect against this and other cybersecurity risks.
Advisories like this name the exposed products attackers go after on your perimeter. See what is exposed on your perimeter with a free passive scan.