Contec Health CMS8000 Patient Monitor (Update A)
Successful exploitation of these vulnerabilities could allow a threat actor to cause a denial-of-service condition, modify firmware with physical access to the device, access a root shell, or employ hard-coded credentials to make configuration changes.
describes (5)
Related CVEs.
| ID | severity | exploitable | ||
|---|---|---|---|---|
| CVE-2022-3027US | medium5.7 | — | ||
The CMS8000 device does not properly control or sanitize the SSID name of a new Wi-Fi access point. A threat actor could create an SSID with a malicious name, including non-standar… | ||||
| CVE-2022-38453US | medium4.4OT elevated | — | ||
Multiple binary application files on the CMS8000 device are compiled with 'not stripped' and 'debug_info' compilation settings. These compiler settings greatly decrease the level o… | ||||
| CVE-2022-38100US | high7.5 | — | ||
The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP req… | ||||
| CVE-2022-38069US | medium6.1OT elevated | — | ||
Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any devi… | ||||
| CVE-2022-36385US | medium6.8 | — | ||
A threat actor with momentary access to the device can plug in a USB drive and perform a malicious firmware update, resulting in permanent changes to device functionality. No authe… | ||||
references
External citations — patches, advisories, analysis.
- ICS Advisory ICSMA-22-244-01 JSONself
- Recommended Practicesexternal
- ICSA Advisory ICSMA-22-244-01 - Web Versionself
- Recommended Practicesexternal
- Recommended Practicesexternal
- Recommended Practicesexternal
- Recommended Practicesexternal
- Recommended Practicesexternal
remediations (5)
- MitigationCMS8000 CONTEC ICU CCU Vital Signs Patient Monitor
Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.
- MitigationCMS8000 CONTEC ICU CCU Vital Signs Patient Monitor
Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.
- MitigationCMS8000 CONTEC ICU CCU Vital Signs Patient Monitor
Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.
- MitigationCMS8000 CONTEC ICU CCU Vital Signs Patient Monitor
Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.
- MitigationCMS8000 CONTEC ICU CCU Vital Signs Patient Monitor
Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.