Fresenius Kabi Agilia Connect Infusion System (Update A)
Successful exploitation of these vulnerabilities in system accessories could allow an attacker to gain access to sensitive information, modify settings or parameters, or perform arbitrary actions as an authenticated user.
describes (13)
Related CVEs.
| ID | severity | exploitable | ||
|---|---|---|---|---|
| CVE-2021-44464US | high8.8OT elevated | — | ||
Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 contains service credentials likely to be common across all instances. An attacker in possession of the password may g… | ||||
| CVE-2021-43355US | critical9.8 | — | ||
Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 allows user input to be validated on the client side without authentication by the server. The server s… | ||||
| CVE-2021-41835US | high7.5OT elevated | — | ||
Fresenius Kabi Agilia Link + version 3.0 does not enforce transport layer encryption. Therefore, transmitted data may be sent in cleartext. Transport layer encryption is offered on… | ||||
| CVE-2021-33848US | medium6.1 | — | ||
Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 is vulnerable to reflected cross-site scripting attacks. An attacker could inject JavaScript in a GET p… | ||||
| CVE-2021-33846US | high7.2OT elevated | — | ||
Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 issues authentication tokens to authenticated users that are signed with a symmetric encryption key. An… | ||||
| CVE-2021-33843US | medium5.3OT elevated | — | ||
Fresenius Kabi Agilia SP MC WiFi vD25 and prior has a default configuration page accessible without authentication. An attacker may use this functionality to change the exposed con… | ||||
| CVE-2021-31562US | critical9.1OT elevated | — | ||
The SSL/TLS configuration of Fresenius Kabi Agilia Link + version 3.0 has serious deficiencies that may allow an attacker to compromise SSL/TLS sessions in different ways. An attac… | ||||
| CVE-2021-23236US | high7.5 | — | ||
Requests may be used to interrupt the normal operation of the device. When exploited, Fresenius Kabi Agilia Link+ version 3.0 must be rebooted via a hard reset triggered by pressin… | ||||
| CVE-2021-23233US | critical9.8OT elevated | — | ||
Sensitive endpoints in Fresenius Kabi Agilia Link+ v3.0 and prior can be accessed without any authentication information such as the session cookie. An attacker can send requests t… | ||||
| CVE-2021-23207US | medium5.5 | — | ||
An attacker with physical access to the host can extract the secrets from the registry and create valid JWT tokens for the Fresenius Kabi Vigilant MasterMed version 2.0.1.3 applica… | ||||
| CVE-2021-23196US | critical9.8 | — | ||
The web application on Agilia Link+ version 3.0 implements authentication and session management mechanisms exclusively on the client-side and does not protect authentication attri… | ||||
| CVE-2021-23195US | medium5.3 | — | ||
Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 has the option for automated indexing (directory listing) activated. When accessing a directory, a web… | ||||
| CVE-2020-35340US | high7.5 | — | ||
A local file inclusion vulnerability in ExpertPDF 9.5.0 through 14.1.0 allows attackers to read the file contents from files that the running ExpertPDF process has access to read. | ||||
references
External citations — patches, advisories, analysis.
remediations (39)
- MitigationAgilia Connect WiFi module of the pumps
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Connect WiFi module of the pumps
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Connect WiFi module of the pumps
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Connect WiFi module of the pumps
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Connect WiFi module of the pumps
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Connect WiFi module of the pumps
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Connect WiFi module of the pumps
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Connect WiFi module of the pumps
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Connect WiFi module of the pumps
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Connect WiFi module of the pumps
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Connect WiFi module of the pumps
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Connect WiFi module of the pumps
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Connect WiFi module of the pumps
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Link+
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Link+
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Link+
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Link+
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Link+
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Link+
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Link+
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Link+
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Link+
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Link+
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Link+
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Link+
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Link+
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Partner maintenance software
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Partner maintenance software
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Partner maintenance software
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Partner maintenance software
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Partner maintenance software
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Partner maintenance software
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Partner maintenance software
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Partner maintenance software
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Partner maintenance software
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Partner maintenance software
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Partner maintenance software
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Partner maintenance software
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.
- MitigationAgilia Partner maintenance software
Fresenius Kabi also identified that early Link+ devices (approximatively 1,200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in users’ installation, Fresenius Kabi recommends users to rely on CISA recommendations below.