ICSMA-21-273-01

icsaUScisa_csaf_ot

Boston Scientific Zoom Latitude

view advisory published 2021-09-30

Successful exploitation of these vulnerabilities may allow an attacker with physical access to the affected device to obtain patient protected health information (PHI), and/or compromise the integrity of the device. The affected device is not network connected and does not contain hardware to be network connected.

01

describes (5)

Related CVEs.

IDseverity exploitable
CVE-2021-38400USmedium6.8—
An attacker with physical access to Boston Scientific Zoom Latitude Model 3120 can remove the hard disk drive or create a specially crafted USB to extract the password hash for bru…
CVE-2021-38398USmedium6.8—
The affected device uses off-the-shelf software components that contain unpatched vulnerabilities. A malicious attacker with physical access to the affected device could exploit th…
CVE-2021-38396USmedium6.8OT elevated—
The programmer installation utility does not perform a cryptographic authenticity or integrity checks of the software on the flash drive. An attacker could leverage this weakness t…
CVE-2021-38394USmedium6.4—
An attacker with physical access to the device can extract the binary that checks for the hardware key and reverse engineer it, which could be used to create a physical duplicate o…
CVE-2021-38392UShigh7.6—
A skilled attacker with physical access to the affected device can gain access to the hard disk drive of the device to change the telemetry region and could use this setting to int…
02

references

External citations — patches, advisories, analysis.

03

remediations (5)

  • FixedZOOM LATITUDE Programmer/Recorder/Monitor (PRM) Model

    Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security, the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120.

  • FixedZOOM LATITUDE Programmer/Recorder/Monitor (PRM) Model

    Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security, the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120.

  • FixedZOOM LATITUDE Programmer/Recorder/Monitor (PRM) Model

    Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security, the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120.

  • FixedZOOM LATITUDE Programmer/Recorder/Monitor (PRM) Model

    Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security, the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120.

  • FixedZOOM LATITUDE Programmer/Recorder/Monitor (PRM) Model

    Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security, the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120.