Hamilton-T1
Successful exploitation of these vulnerabilities could allow attackers with physical access to the device to obtain sensitive information or crash the device being accessed.
describes (3)
Related CVEs.
| ID | severity | exploitable | ||
|---|---|---|---|---|
| CVE-2020-27290US | medium4.3 | — | ||
In Hamilton Medical AG,T1-Ventillator versions 2.2.3 and prior, an information disclosure vulnerability in the ventilator allows attackers with physical access to the configuration… | ||||
| CVE-2020-27282US | medium4.3 | — | ||
In Hamilton Medical AG,T1-Ventillator versions 2.2.3 and prior, an XML validation vulnerability in the ventilator allows privileged attackers with physical access to render the dev… | ||||
| CVE-2020-27278US | medium5.2OT elevated | — | ||
In Hamilton Medical AG,T1-Ventillator versions 2.2.3 and prior, hard-coded credentials in the ventilator allow attackers with physical access to obtain admin privileges for the dev… | ||||
references
External citations — patches, advisories, analysis.
- ICS Advisory ICSMA-21-047-01 JSONself
- Recommended Practicesexternal
- Recommended Practicesexternal
- Recommended Practicesexternal
remediations (3)
- FixedT1 Ventilator
Hamilton Medical recommends users of affected versions update to a version later than T1 Ventilator Version 2.2.3 where possible. Additionally, Hamilton Medical recommends users:
- FixedT1 Ventilator
Hamilton Medical recommends users of affected versions update to a version later than T1 Ventilator Version 2.2.3 where possible. Additionally, Hamilton Medical recommends users:
- FixedT1 Ventilator
Hamilton Medical recommends users of affected versions update to a version later than T1 Ventilator Version 2.2.3 where possible. Additionally, Hamilton Medical recommends users: