Medtronic 2090 Carelink Programmer Vulnerabilities (Update C)
Successful exploitation of these vulnerabilities may allow an attacker with physical access to a 2090 Programmer to obtain per-product credentials to the software deployment network. These credentials grant access to the software deployment network, but access is limited to read-only versions of device software applications.
describes (3)
Related CVEs.
| ID | severity | exploitable | ||
|---|---|---|---|---|
| CVE-2018-10596US | high8.0 | — | ||
Medtronic 2090 CareLink Programmer
uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private networ… | ||||
| CVE-2018-5448US | medium5.7 | — | ||
Medtronic 2090 CareLink Programmer’s software deployment network contains a directory traversal vulnerability that could allow an attacker to read files on the system. | ||||
| CVE-2018-5446US | medium5.3 | — | ||
Medtronic 2090 CareLink Programmer
uses a per-product username and password that is stored in a recoverable format. | ||||
references
External citations — patches, advisories, analysis.
remediations (6)
- Fixed2090 CareLink Programmer
After additional review and risk evaluation of the affected products, Medtronic has disabled the network-based software update mechanism, including both the VPN and the HTTP subservices, as an immediate security mitigation. Users should not attempt to update the affected products over the network as this update mechanism is vulnerable to the attack described in section 4.2.3. Medtronic will continue to implement and deploy increased security protections and mitigations to address the vulnerabilities in this advisory.
- Fixed2090 CareLink Programmer
After additional review and risk evaluation of the affected products, Medtronic has disabled the network-based software update mechanism, including both the VPN and the HTTP subservices, as an immediate security mitigation. Users should not attempt to update the affected products over the network as this update mechanism is vulnerable to the attack described in section 4.2.3. Medtronic will continue to implement and deploy increased security protections and mitigations to address the vulnerabilities in this advisory.
- Fixed2090 CareLink Programmer
After additional review and risk evaluation of the affected products, Medtronic has disabled the network-based software update mechanism, including both the VPN and the HTTP subservices, as an immediate security mitigation. Users should not attempt to update the affected products over the network as this update mechanism is vulnerable to the attack described in section 4.2.3. Medtronic will continue to implement and deploy increased security protections and mitigations to address the vulnerabilities in this advisory.
- Fixed29901 Encore Programmer
After additional review and risk evaluation of the affected products, Medtronic has disabled the network-based software update mechanism, including both the VPN and the HTTP subservices, as an immediate security mitigation. Users should not attempt to update the affected products over the network as this update mechanism is vulnerable to the attack described in section 4.2.3. Medtronic will continue to implement and deploy increased security protections and mitigations to address the vulnerabilities in this advisory.
- Fixed29901 Encore Programmer
After additional review and risk evaluation of the affected products, Medtronic has disabled the network-based software update mechanism, including both the VPN and the HTTP subservices, as an immediate security mitigation. Users should not attempt to update the affected products over the network as this update mechanism is vulnerable to the attack described in section 4.2.3. Medtronic will continue to implement and deploy increased security protections and mitigations to address the vulnerabilities in this advisory.
- Fixed29901 Encore Programmer
After additional review and risk evaluation of the affected products, Medtronic has disabled the network-based software update mechanism, including both the VPN and the HTTP subservices, as an immediate security mitigation. Users should not attempt to update the affected products over the network as this update mechanism is vulnerable to the attack described in section 4.2.3. Medtronic will continue to implement and deploy increased security protections and mitigations to address the vulnerabilities in this advisory.
Advisories like this name the exposed products attackers go after on your perimeter. See what is exposed on your perimeter with a free passive scan.